Product Main

Quick Details

Brand Name: NORDA
Name: Laser Particle counter
Certificate: NORDA
Weight: 5.5 kg
Usage: clean room test
Model Number: ND-6300 SERIES
Place of Origin: China (Mainland)
Flow rate: 1 CFM 28.3L/min,50L/min
name: Laser Particle counter
screen: touch screen

Specifications

Laser particle counter ND-6300(1CFM/ 28.3L/min or 50L/min) is a kind of instrument used to measure the size and quantity of dust particles in unit volume air of clean environment, which can directly detect A and D class clean environment. with features of using diode laser light source, large touch-screen display, small size, light weight, high precision, simple operation, microprocessor control, storing & printing the measurement results and so on, this product is widely used in electronics, optical, chemical, food, cosmetics, health, biological products, aerospace, etc.


[Features

  1. Flow rate:50L /min(50LPM)[ND-6350(T)] or 28.3L/min(28.3LPM)[ND-6330(T)]
  2. Stainless steel housing, easy to surface cleaning and disinfection.
  3. Display six-channel particle sizes and particle concentration (pcs/m3).
  4. Automatic UCL calculation (for the 2010 version of GMP, Euro version of GMP, ISO14644, 16292-2010,etc.)
  5. 95% UCL calculation of particle concentration can be displayed directly (particles / cubic meter, grain / cubic feet)
  6. Touch screen, light weight, small size
  7. Easy to store,print,and U disk output
  8. Easy and simple interface,easily to operate this unit after reading operation manual one time
  9. Connect the compressed air gatherer (model:GC100)(Optional)

[Specification ]                            /   

Overall dim.(mm) 230x 220x 200(wxdxh)
Display Touch screen (English and ChineseVersion ) 
Weight 5.5 kg
Flow rate 50L/min or 28.3L/min 1CFM(optional)
Size channels 0.3,0.5,1,3,5,10(m)
UCL Calculation

95%UCL,display directly (particles/cubicmeter);

(ISO14644-1,Eur GMP,16292-2010,GMP(2010), etc.)

Sample time 0-6000sec
Zero count 10min
Printer Built-in thermal printer (stylus printer optional)
Alarm 100,1000,10000(Class)
Max. Particle concentration 35000 particles/L
Operating Time 4 hours or above
Light source Semiconductor laser(30000 hours)
Date storage 6000 groups
Working condition Temperature:10~35; Humidity:20%~75%RH; Atmosphere Pressure: 86~106Kpa
Optional Trolley for clean room























Laser Particle counter with touch screen 1 CFM model ND6300T series1 CFM 283Lmin50Lmin