Mode and display and operation method:
Mode: A/C, A/C+SIGH,SIMV, SIMV+PSV,PCV, PSV,MMV,PEEP,CPAP
Display: 10.4’ TFT color screen
Operation method: knob operation and touch key
Technical parameter:
Ventilation wave type: squire wave, deceleration wave
Tidal volume: 50~1500ml Breath frequency: 2~60 BPM I: E ratio: 3:1~1:3
Inspiration plateau: 0~50% PEEP: 0~2.0 KPa Trigger sensitivity: -0.1~2.0 KPa, OFF
Oxygen concentration: 21~100% PSV pressure: 0~3.5KPa MV: 1.0~30.0 L/min
Monitor wave and parameters:
Monitor wave: P-t, F-t Tidal volume Breath frequency I:E ratio
PEEP Ppeak MV Oxygen concentration Pmean
Alarm parameters and alarm setting
Tidal volume upper limit: 50~3000 ml Tidal volume lower limit: 0~1000 ml
Breath frequency upper limit: 10~100 BPM Breath frequency lower limit: 2~30 BPM
Peak pressure upper limit: 1.0~6.5 KPa Peak pressure lower limit: 0~2.0 KPa
PEEP upper limit: 0.5~3.0 KPa PEEP lower limit: 0~2.0 KPa
Oxygen concentration upper limit: 30%~100% Oxygen concentration lower limit: 21%~60%
MV upper limit: 2.0~40 L/min MV lower limit: 0.5~20 L/min
Apnea alarm and setting: 3~12 S