Product Main

Specifications

LCD Display Portable Anesthesia Machine with IPPV and SIMV Respiration Mode

 

With IPPV and SIMV respiration mode;

With LCD display, the waveforms showed are: P-t, V-t;

Monitoring tidal volume, minute ventilation, Ppeak, PEEP, respiratory frequency, I:E ratio, and high /lower pressure alarm function, to make sure the operation safety,

60 cmH2O peak pressure protection,

With pediatric bellow, when set small tidal volume for pediatric, the ventilator will automatically change to pediatric mode (no need operation)

International standard connector with Readeagle anesthesia machine and other brands anesthesia machine,

Gas supply: O2, or Air

Applications:

The Ventilator is newly developed portable unit, suitable for respiratory management of full anesthestic condition. Light, small, movable, for patient of more than 10 kgs. It can work together with anesthesitic ventilation system and perform tightly closed, semi-tightly closed and semi-open respiratory management.

 

Specifications:

Display

5.5” SNT display

Waveform

P-t, V-t

Ventilation Volume

50~ 1600ml

Breath frequency

1 ~ 65BPM

I:E ratio

8:1 ~ 1:8

Monitor parameter

 

Tidal volume,

MV,

Ppeak,

PEEP,

Respiratory frequency,

I:E ratio,

Pressure alarm limit

 

High: 20~60cmH2O

 

Low: 0~20cmH2O

Gas supply

O2 or Air

Gas Supply pressure:

O2(Air): 0.25~0.65Mpa

Physical specification

Packing size:33 x 27 x 41 cm

Gross weight: 8.0kgs

Environments

 

Environment Temperature: -9~+35°C

Relative humidity limits: 20%~60%

Atmosphere pressure: 700~1060hPa

Competitive Advantage:

With IPPV and SIMV respiration mode;

With LCD display, the waveforms showed are: P-t, V-t;

Monitoring tidal volume, minute ventilation, Ppeak, PEEP, respiratory frequency, I:E ratio, and high /lower pressure alarm function, to make sure the operation safety,

60 cmH2O peak pressure protection,

With pediatric bellow, when set small tidal volume for pediatric, the ventilator will automatically change to pediatric mode (no need operation)

International standard connector with Readeagle anesthesia machine and other brands anesthesia machine,

Gas supply: O2, or Air