Product Main

Specifications

Handheld Pulse oximeter


Specifications:

1. General


- Monitoring Parameters: SpO2, PR (Pulse Rate)

- Signal sockets: SpO2 socket, adaptor socket

2. Display screen

- Type: Red LED

- Display area: 56mmX36mm

- Size: 120mmX63mmX32mm

- Weight: 300g (not include probes and adaptor)

3. Features

- Audio and visual alarm: The speaker sounds the alarm and the parameter's measurement flashes on the screen.

- Prompting: Icons of battery capacity, pulse sound switch and alarm sound switch.

- Power saving: The monitor will automatic shut down when no monitoring for 3 minutes.

4. Electrical specifications

- Working Voltage: DC6.0V-DC9.0V

- AC power adaptor

- Input: AC100 V-240V, 50/60Hz, 0.7A

- Output: 8V DC, 800mA

- Internal battery

- Type: 2Ahr/6V Ni/MH rechargeable battery

- Run Time:15-hour continuous operation with a new, fully charged battery (environment temperature is 25℃)

- Recharge time: 10 hours

5. Environment

♦ Temperature


- Operation: 0 ~ 45 (°C)

- Transportation and storage: -20 ~ 60 (°C)

♦ Humidity

- Operation: 15% - 95 %( noncondensing)

- Transportation and storage: 10% - 95 % (noncondensing)

♦ Altitude

- Operation: 86 KPa ~106 KPa

- Transportation and storage: 50KPa~106 KPa

6. Parameter Specifications

♦ SpO2


- Patient: Adult, child

- Range: 35 - 99%

- Resolution: 1 %

- Accuracy: ±2 %(70% - 99%),

♦ PR

- Range: 0~250bpm

- Resolution: 1bpm

- Accuracy: ±2bpm