1 ,Innovated rich oxygen air-acetylene flame analysis technique(WFX-110B):
The patented flame analysis technique adopting rich oxygen air-acetylene flame as the substitution for nitrous oxide-acetylene flame for high temperature element analyses, such as Ca, Al, Ba, W, Mo, Ti, V, etc. Flame temeperature is continuously adjustable between 2300-2950℃, which makes it possible to choose the best atomization temperature for different elements. It features easy operation, low analysis cost and wide flame AAS analytical range. Rich oxygen flame will not pollute the environment and is not harmful to human bodies. It's break-through in flame AAS analysis
2, Flame atomization system with flame emission burner
A flame emission burner head can be installed to perform flame emission analysis to Alkali metals as K, Na etc.(WFX-110B/120B)
3, Accurate fully automated control system
3.1 Automatic multi-lamp turret, automatic adjustment of lamp current and optimization of light beam position
3.2 Automatic wavelength scanning and peak picking
3.3 Automatic spectral bandwidth changing
3.4 Automatic ignition
Alarm and automatic protection to fuel gas leakage, abnormal flow, insufficient air pressure and abnormal flame extinction in flame system