Place of Origin: | China (Mainland) |
---|---|
Model Number: | SA808 |
Brand Name: | Utrao |
Type: | Biochemical Analysis System |
Name: | SA808 fully automatic analyzer |
Appearance: | Modern and Exquisite |
Refrigeration system: | Freon-free refrigeration |
System: | sysmex hematology analyzer |
Quality: | Excellent |
Feature: | Biochemistry Analyzer |
Principle: | clinical biochemistry analyzer |
Blood Lipid: | CHO, TG, HDL-C, LDL-C |
Blood Glucose: | GLU |
Monitor: | big-screen color monitor |
Quick Details
Specifications
Application Field:
Liver Function: ALT/GPT, AST/GOT, ALP, T.BIL, D.BIL, TP, ALB
Kidney Function: BUN, Cre, CO2, UA
Blood Lipid: CHO, TG, HDL-C, LDL-C
Blood Glucose: GLU
Myocardial Enzyme: CK, CK-MB, LDH, a-HBDH, r-GT
Ion: Ca, P, Mg, K, Na, CL, etc.
Main Features of SA808 reader
It can automatically optimize testing order, reduce waiting time, and improve testing speed.
Automatic stirring and mixing, makes the mixture of sample and reagent more even, reaction more perfect.
Stand-alone freon-free environmental protectional refrigeration system, refrigerating for 24 hours without a break, guarantee the quality of the reagent.
Flexible and open reagent scheme, one reagent is available to many items.
Free set of calculation item of multi-item participation.
Completed quality-control scheme, different thickness of high, medium, low, support L-J and X-B quality control.
Dynamic real-time reaction curve monitors the disabled reagent timely, prompt too high thickness and make automatic dilution of sample.
Insert emergency at any time, and start emergency testing at first time.
In running, the graphic guide-view system of reagent and sample tray makes it easier to provide the user to look over and master testing information.
Strong database storage technology, hundreds of storage of result, with no influence to running speed.
Complete setup to reference range as different sex and age provides convenience to clinical reference.
Main technical specification of SA808 |
|
Wavelength |
8 filters, 340,380,405,492,510,546,578,630nm, or match as user's demand. |
Temperature control |
25°C, 30°C, 37°C±0.1°C |
Analysis method |
end point, dynamic, fixed time, sample blank, reagent blank, absorbency, double-wavelength, etc. |
Testing speed |
100 samples/hour |
Reagent capacity |
3 reagent tray, 17 reagents/tray, 51 samples for one time (including standard). |
Crossed contamination |
<1% |
Computer |
Pentium computer chip |
Monitor |
big-screen color monitor |
Printer |
Outside high resolution printer |
Keyboard |
Standard computer keyboard |
Exterior interface |
bidirectional RS-232. USB interface |
Light source |
Long life halogen lamp, 12V,20W |
Reading cell |
A continuous flow-cell:Internal volume 33 ul, usual reading volume 500 ul |
temperature control |
37o C± 0.1o |
Reagent capacity |
Three reagent disks, eachdisk can hold 17 different reagents, 51 reagents supported totally |
Sample & reaction plate |
Round rotatingplate, with 45 positions for sample orstandard cups. 37o C thermostatcontrolled. 4 replaceable multi-cell sectors (24cells each). 96 reaction cellstotally. |
Automatic Calibration |
Automatic; 1-45 (optional) orstandard protein-based calibration solution |
Range of absorbency |
0 3 Abs Linearity: <1% |
CV |
<1% |
Power supply |
220V, 50Hz (optional 115V,60Hz) |
Power |
140W |
External Interface |
Two-way RS-232 Dimensions: 370x650x670mm |
Weight |
38Kg |
Note |
24 pin stylus printer, inkjetprinter or laser printer all could be connected |