Place of Origin: | China (Mainland) |
---|---|
Model Number: | MSLDR04 |
Brand Name: | MSL |
Properties: | Medical X-ray Equipments & Accessories |
Quick Details
Specifications
500ma x-ray machine | High Frequency X-ray Radiography System MSLDR04
High Frequency X-ray Radiography MSLDR04 System configuration
High Frequency High Voltage X-ray Generator MSLG50R-A/ MSLO-Ⅳ
X-ray Tube Assembly H1080
Collimator M-38
Digital Detector 1717SCC
DR Machine Frame MSLDR-Ⅰ
Radiography Table MSLS-Ⅲ
Computer and LCD DCSM1F
Synchronous box MSLT-Ⅰ
High Voltage Cable 75kV (length: 8m)
High Frequency High Voltage X-ray Generator MSLG50R-A
MSLG50R-A/ MSLO-Ⅳ operating console
Power AC380V ± 38V
Radiography Tube Voltage 40kV ~ 150kV
Radiography Tube Current 10mA ~ 630mA
Time 2ms ~ 6300ms
Allowed Internal Resistance 380V: 0.5Ω
Allowed Frequency Variation 50Hz/60Hz ± 1Hz
Working Frequency ≥30kHz
X-ray Tube Assembly H1080
Focus 0.6mm/1.2mm
Rotating Speed 3000rpm
Peek Tube Voltage 150kV
Small Focus 20kW
Large Focus 50kW
Collimator M-38
Mode of adjustment Manual
Power 24VAC
Inherent Filtration 1.2mmAL
Digital Detector1717SCC
Scanning Area 17 × 17 inches
Spatial Resolution not less than 3.6lp/mm
Total Pixel Matrix 3328(h) × 3328(v)
Pixel Pitch ≤127μm
A/D 16Bit
Detector Imaging Time not more than 4S
DR Machine Frame MSLDR-Ⅰ
Up and Down Moving Range of UC arm with the Column 700mm ~ 1700mm
Rotating Range of UC arm along with the Column 0°~ 105°
Up and Down Moving Range of Detector 700mm ~ 1700mm
Moving Range of X-ray Tube Assembly 0 ~ 800mm±50mm
SID 1000mm ~ 1800mm
Radiography Table MSLS-Ⅲ
Tabletop Size (Length × Width × Height) 2020mm × 690mm × 680mm
Mode of Operation Manual
Load-bearing Less than 200kg
Computer and LCD DCSM
CPU ≥2.8GHz
Memory ≥4GB
Disk Capacity ≥500GB
Medical LCD ≥19"