Features
Single channel with measurement report
Build in rechargeable Ni-MH battery
Chinese/English language
Simultaneous 12 leads
320X240 graphic LCD, 3.8 inch screen
Paper size: 50mmX20/30m
Built in RS232 interface (USB is optional)
8 Patients save/copy/communication
Specification
Lead standard 12 leads
lead acquisition synchronously 12 leads
Data acquisition 18bit/1000Hz
Record mode Manual mode; Auto 1; Auto 2; Auto 3
Filter AC, EMG, Drift Filter
CMRR 100dB(with AC filter)
Input circuit Floating; Protection circuit against Defibrillator effect
Input circuit current ≤0.1μA
Input Impedance ≥2.5MΩ
Patient current leakage <10μA
Calibrating voltage 1mV?%
Voltage Tolerance ?00mV
Time constant >3.2s
Frequency response 0.05 ~150Hz
Noisy level ≤15μVp-p
Sensitivity valve ≤20μV
Inter-Channel Interference ≤0.5mm
Paper speed 25, 50mm/s(?%)
Sensitivity AUTO, 2.5, 5, 10, 20mm/mV
Recording mode Thermal Array printing
Paper specification 50mmx20m, roll
LCD display 320X240 graphic LCD, display status & waveform
Safety classification IEC60601-1 class I, type CF
Power supply AC : 110V/220V/85~265V, 50/60Hz?Hz; 30VA
DC : 12V(1500mAh) rechargeable Ni-MH battery