Standard parameter: ECG, RESP, NIBP, SpO2, TEMP, PR.

12.1" large-screen color TFT display

Built-in high-capacity battery, up to 4 hours working time

7-lead ECG waveforms display ,big fonts observational interface

Reviewable ECG waveforms, Maximum 240 hours reviewing graphic trends

Digital SpO2 technique, precise measurement for weak perfusion and finger shaking

Arrhythmia and S-T segment analysis and pacemaker detection

Protection against interference from defibrillation and electrosurgical Unit

Applicable to ICU, CCU, operation, adult, pediatrics and neonatal patient

 

ECG

l          Full lead:     I, II, III, AVL, AVR, AVF and V
l          Gain       5mm/mV, 10mm/mV, 15mm/mV, 20mm/mV and 25mm/mV
l          Scanning speed: Changeable with three grades of 12.5mm/s, 25mm/s and 50mm/s
l          Band width          0.05Hz-100Hz(3dB)
l          Heart rate:
Scope:          15 times/minute to 300 times/minutes
Accuracy:     ±1 time/minute
l          Variation of heart beat        LORENZE drawing: Analysis on heart beat data for the current 5 minutes with histogram during RR period.
l          Arrhythmia :Twenty kinds of common abnormal ECG can be analyzed, with ECG data of 1000 times of arrhythmia kept and with 10 seconds kept for each time.   
Accuracy:     Error rate < 20%        
Undetected rate < 20%
Analysis function should be started when using the above two kinds of function.
l          ST section
Precision :     < ±0.05mV

SpO2

l          Detection scope: 0%-100%
                 ±2%    70%-89%

NIBP

l        Theory         Electric oscillator
l        Cuff gassing time   less than 15 seconds
l        First-time gassing pressure: dynamic state 24kPa (180 mmHg) 
Manual selection: 9.3kPa (70 mmHg),13.3kPa (100 mmHg),16.0kPa
(120 mmHg)
18.6kPa (140 mmHg),20.0kPa (150 mmHg),21.3kPa (160 mmHg),
l        Automatic detection period: 1 minute to 240 minutes
Stop size:1 min(1 min -10min), 5min(10 min -30min), 10min(30 min -90min) and 30min(90 min -240min)
l        Detection scope:
Systolic blood pressure:     4kPa -34kPa(30mmHg -255mmHg)
Diastolic pressure              2kPa -29.3kPa(15mmHg -220mmHg)
Mean pressure            2.7kPa~31.3kPa (20mmHg~235mmHg)
l        Detection accuracy
Static pressure: ±3mmHg
Pulse rate:            ±2%

Resp

l        Chest impedance method or concentration of carbon dioxide at end of respiration
l        Scope of chest impedance method   0 time/minute – 100 minutes/minute
l        Precision of chest impedance method       1 time/minute
l        Detection scope of carbon oxygen at end of respiration 3mmHg-75mmHg; 0.4%vol-9.9%vol
l        Detection scope of respiration rate: 2 times/minute ->60 times/minute  (including baby)

Temp

l        Single/double path
l        Type:     Vascular/surface temperature
l        Scope:          20℃~50℃
l        Accuracy             ±0.2℃

Invasive blood pressure (selected mating)

l        Single/double path
l        Type: direct and continuous detection
   Scope:            -6.67kPa(-50mmHg)—46.67kPa(350mmHg)
l        Precision:        below 13.33kPa(100mmHg) ±0.27kPa(2mmHg)  13.33kPa(100mmHg) above ±2%
l        Frequency response:    DC-12Hz
l        Input resistance: not less than 500kΩ
l        Sensitivity of sensor: 5μV/V/mmHg
l        Motivation voltage:    5V (DC)
l        Accuracy of sensor: ±2%
l        Isolation voltage: 4 kV