Product Main

Specifications

2. Main specification
(1).Impact energy:   300J
(2).Dial graduation range and value:
Range of energy
0~300J
0~150J
Per graduation value

2J

1J

(3). Pendulum moment:
Pendulum impact energy

300J

150J

Pendulum moment
160.7695N.m

80.3848N.m

(4)Pendulum preparing angle:                       150o
(5)Distance between pendulum center and impact point:  750mm
(6)Impact velocity:                                5.2m/s
(7)Span of specimen seat:                           40mm
(8)End face radius of specimen seat:                  1~1.5mm
(9)Tilt angle of specimen seat support:                 0o
(10).Radius of impact knife:                         2~2.5mm
(11).Angle of impact knife:                         30o±1o
(12).Thickness of impact knife:                      16mm
(13.specifications of specimen:                       10 x 10 x 55mm
(14).Net weight:                                   About 450kg
(15).Dimension:                                   2124╳600╳1340mm
(16).Power supply:                                 1 phases, 60Hz,220V
(17).Power of motor:                                180W
(18).Working condition:
a)         Ambient temperature: 10-35℃
b)        Relative humidity: ≤85%
c)        Around it without corrosive medium
d)        Installing on the concrete groundwork, which thickness is not less than 150mm. or fixing on the base, which weight is over 880kg. 
e)         Levelness of installing datum plane of machine base is 0.2/1000.
 If you have other questions, could you telephone us or go to www.jnxdy.com ?
JB-300B Pendulum Impact Testing Machine is developed as per GB/T3808-2002 <Inspection of Pendulum impact testing machine>, and can finish impact test according to GB/T229-1994 <Method of Metal Charpy Gap Impact Test>. It can realize two kinds of impact energy 300J and 150J. The section area of specimen is 10 x 10mm. mainly use to test larger impact toughness ferrous metal, such as: steel and alloy material.
 
Its principle is to calculate difference of the energy before and after pendulum impacting, the difference is energy which absorbed by specimen. The machine can be operated by semiautomatic, easy to operate, and high efficiency. Furthermore, it can swing up the pendulum automatically by using residual energy. It is especially suitable for continuous test.