Product Main

Specifications

Features
? All Digitized;
? Select simultaneously acquisition of 12-lead ECG signals;
? Powerful anti-interference with variety of filtering methods: DFT, AC,EMG, Lowpass Filter, recorder Filter. Ensure the waveform clearly, accurately and nondeformed;
? Single channel, 1+R(Rhythm)channel, 60 seconds compressible for any channel;
? Double channal synchronous printing selected. Save the half recording paper and the half recording time, and the diagnosis accuracy is double more by two channels compared and measured;
? 60 seconds simultaneous record selected(60 seconds compressible record need 10cm paper but for ordinal Single channel machine need 120cm paper.) to acquire long ECG signal which can diagnose arrhythmia and occasional premature beat etc.;
? Lead (un)connected indication fuction: The machine provide indicating light of connected points by chart, and each indicating light shows the corresponding lead connecting condition. 
During the operating, if any lead off, the machine will automatically stop paper printing and the indicating light indicates the lead which is off in corresponding parts. Then works automatically after the leads ready. So it solve the problem of printing linear when a lead failed, which guides doctors operating extremely convenient, time-safing and avoids wasting of paper;
? V1,V3,V5 chest lead for physical mode can record the patient ECG in 15 seconds/each. Easy for physical;
? Automatic analysis;
? Auto/ Manual operation mode;
? The base line autoregulation, increment autoregulation and automatic lead transferability available in automatical mode;
? Manual shift control available in Manual mode;
? SAMSUNG original integration printer, with strong sealing, high reliability and long life;
? AC/ DC supplied and built in rechargeable battery suporting up to 170mins after complete charging;
? USB Interface transferrd supported optional;
? SD card data storage supported optional; 
? ECG workstations supported optional .
Main Specification
Lead                         Standard 12 leads
Operating Mode:              Automatic / Manual
Input Impedance              ≥50MΩ
Input Circuit Current:       ≤ 50nA
Time Constant:               > 3.2 Sec
Paper specification          50mm x 20m
Recording mode               Thermal Array printing
Sensitivity Option:          Automatic
Safety classification        IEC60601-1 class I, type CF
Power Condition:             AC: 220V±10%, 50Hz/60Hz
                             DC: 12V Rechargeable battery & Circuitry