Product Main

Specifications

Content:
Pulse oximeter, Data line, Power adaptor, Software, Pouch, Manual
 
Function:
Measuring Sp02 and PR accurately in hospital, home healthcar, oxygenbar, community medical centre, alpine area , sports healthcare etc.
 
Main Performance :
A. SpO2 value display
B. Pulse rate value display , bar graph display
C. Pulse waveform display
D. Display Mode : True color OLED with 4 direction display
E. Low-voltage indication: color OLED low-voltage indicator appears before working abnormally which is due to low-voltage
F. Automatically power off  function: when the device is under the state of measuring interface(without external probe), it will automatically power off within 5 seconds if the finger falls out of the probe.
G.  The display mode can be changed.
H.  Screen brightness can be changed
I.  A pulse rate sound indication,
J.  With visual and audio alarm function (alarm setting is 35~99%)
K.  With SpO2 value and pulse rate value of storage, the storage date can be uploaded to computers
L.  It can be connected with an external oximeter probe
M.  Real-time date can be transmitted to computers.
N. USB and wireless interface to PC
O.  Synchronous display on the oximeter and computer
P. Inner flash memoery for more than 30 hours date storage
Q.  Advanced software for sleep study and monitoring
R.  Power supply 3.7V rechargeable Lithium polymer battery
S.  Charger: USB charger like Apple Ipod and AC charger
 
Main Parameters:
A. Measurement of SpO2
  Measurement Range: 0-100%
  Accuracy: 70~100% , ± 2% ;0~69% , unspecified
B. Measurement of pulse rate
  Measurement Range: 25-250bpm
  Accuracy: ± 2bpm or ±2% (select larger)
C. Resolution
SpO2 : 1%, Pulse rate: 1bpm
D. Measurement Performance in Weak Filling Condition:
SpO2 and pulse rate can be shown correctly when pulse-filling ratio is 0.4% . SpO2 error is ±4%, pulse rate error is ±2bpm or ±2%(select larger) .
E. Pixels: 160*128, 256K color
F. Resistance to sorrounding light :
The deviation between the value measured in the condition of man-made light or indoor natural light and that of darkromm is less that  ±1% .