Product Main

Specifications

Feature:

- stable output by the inverter system

- full AEC mode (auto kV)

- auto release compression after exposure

- high performance dual focus system

- highly sensitive fiber grid bucky device

- 19 step density control

- 2 precise collimation formats

 

Intelligent automatic Exposure Control (AEC)

With the automatic exposure control system, it's possible to produce images with reliable intensity suitable for any film screen, or method of radiography.

Furthermore, it greatly enhances the convenience of radiography by embedding the full-AEC function which is capable of utilizing the auto kv

Specification:

 

Generator:

 

- Type: high frequency inverter

- Radiographic ratings: large focal point 22-35kv / 1-600ma ,

  small focal point 22-35kv / 1-100ma

 

X-ray Tube

 

- Focal spot size: dual focus 0.3 and 0.1mm

- rotating anode: molybdenum(Mo)/300KHU

- Port material: beryllium(Be)

- Filtration: Mo

 

Radiographic stand

 

- C-ARM: vertical 560-1340mm , rotation right 180 / left 160 degree

- SID: 600mm fixed distance

- Compression: Manual and motorized(max 8kg) , automatic or manual release

 

Bucky Device:

 

- Gassette size: 18 x 24cm

- GRID: reciprocating, 4:191 line/inch

 

Automatic Exposure Control

 

- Type: solid-state detector , microprocessor control

- Mode: full/semi/manual

- Density adjustment: 19 density steps

 

Line voltage

 

- 200~220 Vac, 50/60Hz, singe phase