Product Main

Specifications

1. Work condition
1.1 Power supply: AC 90V~AC 242V, 50HZ
1.2 Built-in battery: DC 12V 1800mAh rechargeable Ni-MH battery, the battery can be used continuously for 2 hours after full charged (24hours) under 25ml/h flow rate.
1.3 Environmental condition:
 temperature +5℃~43℃           
relative humidity:  ≤90%RH (Under without condensation status)
1.4 Atmospheric pressure: 550hPa~1060hPa
1.5 Storage condition:
  temperature: -40℃~+55℃
relative humidity ≤95%RH (Under without condensation status)
 
2. Basic parameter                                    
2.1 Outline size (mm): 140(L)×150(W)×220(H)
2.2 Weight: about 1.8kg
2.3 Power consumption: ≤10VA
2.4 Flow rate options: 1ml/h~1200ml/h, in 1ml/h increment
2.5 Infusion total amount setting: 1ml~9999ml, in 1ml increment
2.6 Occlusion pressure level:
High:   825mmHg ± 150 mmHg (110kPa±20kPa)
Middle: 600mmHg ± 225mmHg (80kPa±30kPa)
Low:   450mmHg ± 150 mmHg (60kPa±20kPa)
2.7 Flow rate error: ±5%
2.8 Alarm: occlusion, finish, low battery, standby (no operation), air in line, door open, motor troubles, AC power disconnected, starting up by battery
2.9 Function: self-test, infusion amount setting, flow rate and infused volume display, power indication, safe flow rate limit, data preserves, KVO (Keep Vein Open) function