Features
◆ End point, kinetic, fixed time, mutistandard, bichromatic etc
◆ Filter wavelengths: 340,405,492,510,546,578,620nm, 2 more open filter positions, others are available on request
◆ With 20 incubating positions
◆ Large memory to store 200 test programs and 1000 testing results
◆ Excellent Q.C function Q.C. chart can be stored, displayed and printed.
Technical Specifications
● Reading Cuvette: Both through cell and direct reading cuvette
● Incubating Positions: 20 incubation positions
● Photometric System:
Light Source: 6V 10W long-life halogen lamp
Filters: 340,405,492,510,546,578,620nm, 2 more open filter positions; ±2nm pass-band
Wavelength Accuracy: ±2 nm
● Measuring System:
Measuring Range: 0-2.500 O.D
Photometric Linearity: ±2% from 0 to 2.000 O.D.
Photometric Accuracy: ±2% from 0 to 2.000 O.D.
Drift: less than 0.005 O.D.
Carry Over: ≤1%
● Incubator Temperature control: 25℃,30℃,37℃, Precision:±0.1℃
● Aspiration volume: 500ul
● Reagent Volume: 200-500 ul
● Display: Back-illuminated LCD
● Printer: Built-in thermal printer, 57mm width paper
● Power Supply: 110V 60HZ / 220V 50HZ, 60VA
● Dimensions: 34cm*38cm*18cm
● Weight: 8.5kg