Product Main

Specifications

Model
Electrical Safety Comprehensive Tester AN9651B(F) (6 in 1)
Display Mode
240×128 LCD Large Screen
Isolated Transformer Capacity
Single phase: 2000W/6000W
Over-current Protection Limit
When current 30A, exceed 5s protection
Alarm Mode
Alarm with sound and light
Standard Interfaces
RS232 interface, Alarm Light interface, Remote Control interface, AN97 Frequency Converter Power Supply interface
Interfaces Optional
RS485 interface, Printer interface, Foot Pedal interface, On-off interface
Buzzer Alarm Sound
Can ON and OFF
Test Group
15 Group, 7 test steps/group
Ground Bond Test
No-load Output Voltage
< 12V
Current Presetting Range
AC 3A~30A 50Hz/60Hz 1A/step
Current Accuracy
±(1. 5% of setting+0. 40A)
Current Stability
< 0. 4% (10A, 100mΩ)
Resistance Range/Accuracy
10~600mΩ  ±(1. 5% of reading+1. 5% of Range)
Resistance Upp-Limit Range
3A≤Current≤10A: 10~600mΩ; 11A≤Current≤25A: 10~300mΩ;
26A≤Current≤30A: 10~200mΩ   1mΩ/step
Resistance Low-Limit Range
0mΩ
Compensation Resistance Range
0~200mΩ, Automatic Measurement, ON/OFF
Voltage Range/Accuracy
30~7500mV ±(1. 5% of reading+1. 5% of Range)
Voltage Upp-Limit Range
30~7500mV  1mV/step
Voltage Low-Limit Range
0mV
Compensation Voltage Range
0~1000mV, Automatic Measurement, ON/OFF
Test Time Range/Resolution
1~999s 1s/step
Insulation Resistance Test
Rated Output Load/Current
2VA(1000V/2mA)  2mA
Output Voltage Range/Resolution
DC 500V/1000V < 1% (From no load to full load)
Output Voltage Ripple
< 1% (Pure Resistive 1000V/1mA)
Output Voltage Accuracy/Resolution
±(3% of setting+3V)  5V
Resistance Range/Accuracy
≤200MΩ: ±(3% of reading+3 counts);
> 200MΩ: ±(8% of reading+8 counts)
Resistance Upp-Limit Range
2000MΩ
Resistance Low-Limit Range
0. 3~2000MΩ  0. 1MΩ/step(≤100MΩ), 1MΩ/step(> 100MΩ)
Test Time Range/Accuracy
1~999s 1s/step
AC Hipot Test
Rated Output Load
500VA(AC: 5000V/100mA)
Output Voltage Range/Freq/Resolution
500~5000V 50Hz/60Hz < 3%(From no load to full load)
Output Voltage Setting Range/Accuracy
Static: AC, 500~5000V; Dynamic: AC, 500~2500V;
±(3% of setting+10V)
Breakdown Current Meas. Accuracy
±(3% of reading+0. 03mA)
Breakdown Current Alarm Preset Range
Upp-limit: 0. 10-100. 00mA; Low-limit: 0. 03-10. 00mA; 0. 01mA/step
Compensation Current Range
0. 00-10. 00mA, Automatic Measurement, ON/OFF
Test Time Range
1~999s  1s/step
Rise Time Range
0~100s  1s/step
Leakage Current Test
Output Voltage Range
1. 06 times of input voltage(output by isolated transformer)
Leakage Mode Setting
Static, Dynamic
Output Voltage Meas. Range/Accuracy
AC 60~280V/45~65Hz; ±(0. 5% of reading+2 counts)
Leakage Current Setting Range
Upp-limit: 0~9. 999mA 0. 001mA; Low-limit: 0mA
Compensation Current Setting Range
0~1000uA, Automatic Measurement, ON/OFF
Leakage Current Meas. Range/Accuracy
0. 050mA~9. 999mA/
DC~10kHz: ±(3% of reading+10uA);
10k~1MHz: ±(5% of reading+50uA)
Test Time Range
1~999s 1s/step
Power Test
Output Voltage Range
1. 00 times of input voltage(output by isolated transformer)
Power Setting Range/Resolution
Upp-limit: 6~6000W Low-limit: 0~6000W 1W/step
Output Voltage Meas. Range/Accuracy
AC 60~280V; 45~65Hz: ±(0. 5% of reading+2 counts)
Current Meas. Range/Accuracy
AC 0. 030~3. 999A, 4. 00~25. 00A;
45~65Hz: ±(0. 5% of reading+0. 020A)
Real-Power Meas. Range/Accuracy
10. 0~199. 9W: ±(0. 5% of reading+5W)
200~6000W: ±(0. 5% of reading+30W)
Power Factor Meas. Range/Accuracy
0. 100~1. 000; ±(2% of reading+2 counts)
Test Time Range
1~999s, 1s/step
Low Voltage Start Test
Output Voltage Range
1. 00 times of input voltage(output by isolated transformer)
Current Meas. Range/Resolution
Upp-limit: 0. 30~25. 00A Low-limit: 0. 00~25. 00A    0. 1A
Output Voltage Meas. Range/Accuracy
AC 60~280V; 45~65Hz: ±(0. 5% of reading+2 counts)
Current Meas. Range/Accuracy
AC 0. 030~3. 999A, 4. 00~25. 00A
45~65Hz: ±(0. 5% of reading+0. 010A)
Test Time Range
1~999s 1s/step