These series sterilizers are a kind of pulse-vacuum automatic instant sterilization equipment which adopts steam as sterilization medium. They comply with Grade-B requirements of European standard and suitable for sterilization of all wrapped or non-wrapped, solid, hollow load products type A, hollow load products type B and medical device with built-in pipe.
Technology advantage:
1. 16-bit micro processor imported from USA, the system is powerful and reliable;
2. Automatic water-filling, pulsing, heating, sterilization, drying.
3. Touch key and menu control, easy operation;
4. 3 fixed sterilization program and one sterilization program with one adjustable parameter.
5. Involve BD test, helix test and Vacuum test procedures; Air leakage
rate signified by a pressure change shall be less than 0, 13 kPa/min.
6. Builted-in vacuum pump, pulse vacuum 3-6 times can reach -0. 09MPa(-0. 9Bar), to ensure effective removal of the air so as to get the optimum saturate steam.
7. High penetration ensure the high steam penetrate into the dressing, disinfect bags and dental hand piece intervals quickly, so as to ensure sterilization thoroughly.
8. Unique pulse vacuum drying procedure, the residue moisture shall be less than 0. 2%.
9. LCD real-time displays status, temperature, pressure and time.
10. Perfect fault check system and display the fault information;
11. Equipped with mini printer to document the performed cycle
12. The beep will sound once the sterilization cycle complete.
Safety feature
Heater protect: the system will cut off its power supply once the temperature exceed the preset one, and show the alarm information.
Anti-dry heating: the system will cut off its power supply once the water level is lower than the heating pipe, and show the alarm information.
Safety interlocking door: electronic lock to ensure the sterilizer can begin to work only if the door is closed properly. The door can’t be opened when the chamber is pressurized, or the power supply is not connected.
Over-temperature and over-pressure protect: When the inner temperature or pressure exceed the limit, the system will start up procedures to lower temperature and release pressure.
Automatic release safety valve: When the chamber inside pressure exceeds preset value, the safety valve will release pressure automatically.
Electricity protect: Designed protect function in case of short circuit and leakage current.