Product Main

Specifications

II-1000 Series Infant Incubator
 
? Standrad configuration: High performance microcomputer control; touch film key operation; constant temperature cover with inclined plane; two display screens for set temp, box inner temp; >37℃ set-up; set-off memory; warning for: set-off, fan, sensor, over box inner temp. & temp. deviation; warning silencing function, self-test; wall cupboard and shelf and RS-232 or 485 connector.
 
II-1000YC
? Standard configuration + Infant bed adjusted with stepless.
 
II-1000CM
? Standard configuration + Side door for operation, such as injecting to infant scalp.
 
II-1000SC
? Standard configuration + Dual air curtains.
 
II-1000GL
? Standard configuration + Phototherapy(blue light)unit, LED display timer, display treatment time and total time.
 
II-1000SGL
? Standard configuration + Two phototherapy(blue light)unit, LED display timer, display treatment time and total time.
 
Specifications:
? Power Supply: AC 110/220V±10%, 50/60Hz±2%
? Heating Power: ≤400VA
? Convenience Outlets: AC 110/220V±10%, 50/60Hz±2%
? With respect to IEC601-1, this equipment is Class I, Type B
? Continuous operation with short time loading(VHA)
 
Alarms:
1. Power Alarm
? Power failure: Activates if primary power to the incubator fails or the power code is accidentally disconnected from the wall receptacle.

2. Fan Alarm
? Fan failure: Activate if fan electromotor failure or fan power failure, or fan rotate speed below 1,000 turn/min because of outside force.

3. Sensor Alarm
? Air sensor failed: Air sensor malfunction.
? Sensor failure: Air sensor not connected to the incubator.

4. Deviation Alarm
? High air temperature: Activated if displayed air temperature is 1.5℃ above set point.
? Low air temperature: Activated if displayed air temperature is 2.5℃ below set point.

5. Over Temperature Alarm
? Over air temperature: Activated if incubator temperature arrived 38℃ for set temps <37℃ in Air Mode; arrived 40℃ for set temps >37℃ in Air Mode.

Silence function of "▲/Silence Complex Key”
The "▲/Silence Complex Key” can make following alarm silenced for 15 minutes.
? Deviation Alarm
? Sensor Alarm
? Fan Alarm