Product Main

Specifications

Fischer  & Porter Corporation has the most advanced vortex flowmeter in the world. It may carry out the flow measuring, detection and controlling for various liquids, gases, high/low temperature superheated steam and saturated steam.

Characteristics

1) Shock resistance type: The vortex flowmeter of Fischer & Porter Corporation adopts 4- piece piezoelectric crystals for measuring. Two of them are used to detect the flow of fluid. The other two are used to detect the vibration signal of pipe.

2) Temperature and pressure compensation

3)  Intelligent type: The meter parameters and meter range may be changed by three manual buttons, magnet bar or manual controller on site.

4) Output mode: 4-20mA current signal output/ pulse signal and contact signal.

 

 
Meter Model VT / VR4000 VTC / VRC4000
Connecting Forms Piped Plug-in
Accuracy liquid 1.0%/0.75% 2.0%
Gases/ steam 1.0% 2.0%
Reproducibility ≤± 0.2%( of rate ) ≤± 0.2%( of rate )
Allowable viscosity of measured medium ≤ 7.5mPs ≤ 7.5mPs
Typical flow Range 1:20 1:8 - 1:30
Sensor Flange connection DIN DN15-DN600
Wafer connection DIN DN15-DN300
Insertion type Diameter of Pipeline > DN400
Fluid Temperature Standard -55oC ~ 280oC -55oC ~ 280oC
High temperature -55oC ~ 400oC -55oC ~ 400
Pressure range Standard ≤ 100bar ≤ 25bar
Special According to demands of users According to demands of users
Material

SS304/SS304L/SS316/SS316L
According to demands of users

SS304/SS304L/SS316/SS316L
According to demands of users
Converter Supply Power 14-28V DC Battery-power 14-28V DC Battery-power
Cable Connection M20x1.5, 1/2'' NPT M20x1.5, 1/2'' NPT
Display LCD display LCD display
Communication HART-Protocol
Profibus-DP field bus
FF field bus
HART-Protocol
Profibus-DP filed bus
FF field bus
Ex-Design Exia Exia ll CT6 Exia ll CT6
Exd Exd ll CT6 Exd ll CT6
Protection class standard IP67 IP67
Diving
IP68 IP68
Ambient temperature -40oC ~ 55oC / -55oC ~ 70oC -40oC ~ 55oC / -55oC ~ 70oC
materials Cast aluminum Cast aluminum
Power consumption 1W 1W