Detail Discription
ECG (ECG-1200)
Specification
* Input Circuit : Floating; Protection against Defibrillator effect
* Lead: Standard 12 leads, Lead change automatically
* Patient Current Leakage: < 10 μ A
* Input Impedance: ≥50 ΜΩ
* Calibrating Voltage: 1mV ±3%
* A/D Conversion: 12 bit
* Display: 320×240LCD
* Frequency Response: 0. 05Hz-150Hz
* Time Constant: > 3. 2seconds
* CMRR: > 80d> 100dB(with filter)
* Paper Speed: 25, 50mm/s±3%
* Recording Paper: 210mm×30m, high-speed roll
* Input Circuit Current: ≤50nA
* Safety Standard: IEC class I, type CF
* Sensitivity: 1/2, 1, 2(cm/mV)
* Skin Voltage Tolerance: ±300mV
* Noise Level: < 15 μVp-p
* Weight: 5. 1kg
Feature:
* Simultaneous 12 leads acquisition and printing
* Automatic measurement and interpretation; R-R analysis
* 12 leads simultaneous display on large backlight LCD 320×240
* Digital design and filter ; Thermal array printer
* Recording on A4 size paper
* Baseline auto-adjustable; Rhythm lead selectable
* DC/AC autoexchange, built-in rechargeable battery. Power supports continuously working for more than 50 patients
* 16 pieces of unzipped ECG date and waveforms are stored for copy and re-call
* Inner handhold designed for portability