Product Main

Specifications

AJ-2102A Anesthesia Machine

*This machine uses in respiratory control , respiratory intensive care , 5.1 inch high clear display, and real-time reflection respiratory system condition.
*The high-precision evaporator has the stable output concentration with pressure, temperature, and flow automatic compensation; Choose one among Enflurane, isoflurane and sevoflurane for a matching.
*Reliable gas Power transmission system;
*Reasonable economic and multiple technology configuration, the user s may choose flexibly;
*Modular product design According to customer’s needs upgrade.

Main Technical Parameters
Part of Anesthesia Machine
*Console high-strength engineering plastic rack,light,beautiful and corrosion resistance
*Scope adults, children
*Gas source O2,N2O:0.3~0.5MPa
*Flow meter O2: 0.1~10L/min
N2O: o.1~10L/min
*O2,N2O linkage and nitrous oxide Truncation device
When use nitrous oxide, oxygen concentration>25%; when the oxygen pressure<0.2Kpa, the flow of nitrous oxide would be cut off.
*Rapid oxygen flow 35~75L/min
*Vaporizer
It has the function of pressure, temperature, and flow automatic compensation. The regulation range of evaporator concentration is 0~5vol%. Choose one among Enflurane, Isoflurane and Sevoflurane for a matching according to the demands of customer.
*Low oxygen pressure alarm
It will be alarmed when the oxygen pressure < 0.2MPa
*Respiratory circuit
Work mode all close. semi-close. semi-open pressure release 1~4KPa
*Bellows breathing
Adults and children bellows tidal volume range:0~1500ml

Part of Ventilator
*Display mode 5.1-inch high clear display,
*Ventilation mode
IPPV SIPPV PLV PEEP MANU (monitor the tidal volume. ventilation. respiratory rate. ratio. peak pressure when use hand.)
*Ventilation parameters
tidal volume 50~1500ml
frequency 6~60bpm
I:E 2:1~1:4
Trigger sensitivity -1.0~-0.2Kpa
PEEP 0.2~1.0Kpa
pressure control 1~6.0Kpa
*Ventilation monitoring parameters
tidal volume. ventilation. IPPV ratio. total respiratory rate. breath /expiratory ratio. peak pressure . pressure control. airway pressure wave. breath pressure trigger. PEEP. minimum pressure .
*Monitoring oxygen concentration 21%~100%(choose and match)
*Security Alert System
Oxygen Concentration alarm (choose and match)
upper limit setting range 50%~100%
low limit setting range 15%~50%
Airway pressure alarm
upper limit setting range 1.5~6.0Kpa
low limit setting range 0.2~3.0Kpa
Alarm minute ventilation
upper limit setting range 2~30L/min
low limit setting range 1~12L/min
Asphyxia alarm The machine will give sound and light alarm if there is no tidal volume input in 15s
Power failure alarm
*Power AC 220V±10% 50Hz±1Hz UPS. Battery. (choose and match)