ECG (ECG-100G)
Features
■ High-resolution thermal printing array system
■ Frequency response is as high as 150Hz
■ Capable of printing continuously on channel trace and annotations
including lead mark and parameter such as sensitivity, paper speed and filter
operation status
■ High resolution digital filter inhibits baseline drift without affecting ECGwaveform
■ One-touch operation
■ Concept of floating input circuit meets IEC safety standard
■ Rechargeable battery supporting more than 50 patient examinations
■ Lead Standard: 12 leads
Specifications
Input Circuit | Floating; Protection against Defibrillator effect |
Lead | Standard 12 Leads, Lead change automatically |
Patient Current Leakage | <10μA |
Input Impedance | ≥10ΜΩ |
Calibrating Voltage | 1mV±3% |
A/D Conversion | 12 bit |
Frequency Response | 0.05Hz~150Hz (IEC) |
Time Constant | >3.2seconds |
CMRR |
>80dB
>100dB (with filter) |
EMG Filter | 35Hz (-3dB) / 25Hz (-3dB) |
Sensitivity | 1/2, 1, 2 (cm/mV), Conversion Deviation≤5% |
Recording System | Thermal-array
8 dots/mm ( vertical )
16 dots/mm ( horizontal, 25mm/sec ) |
Paper Speed | 25, 50mm/s±3% |
Recording Paper | 50mm,20m, high-speed roll |
Input Circuit Current | ≤50nA |
Safety Standard | IEC class I, type CF |
Power Supply |
AC: 220V / 110V, 50/60Hz
DC: 12V rechargeable battery |
Skin Voltage Tolerance | ±300mV |
Noise Level | <15uVp-p |
Dimensions | 310mm×230mm×70mm |
Dimension of Packing | 400mm×280mm×160mm |
Weight | 2.75kg |