| Model | UV-2100PC | UV-2400PC |
| Optical Specifications | ||
| Wavelength Range | 200 - 1000 nm | |
| Spectral Bandwidth | 4 nm | 2 nm |
| Monochromator | Single beam C-T type, grating 1200 lines/mm | |
| Wavelength Accuracy | ± 1.0 nm | ± 0.5nm |
| Wavelength Repeatability | ± 0.5 nm | ± 0.2 nm |
| Wavelength Calibration | Automatic at start | |
| Light Source | Tungsten Halogen Lamp Deuterium Lamp | |
| Detector | Solid Silicon Photodiode | |
| Photometric Readout | 20 X 2 LCD | |
| Photometric Range | 0 - 125%T -0.097 - 2.70A 0 - 1999C (0-1999F) | |
| Photometric Accuracy | ±0.004A(0.5-1A), ±0.002A(0-0.5A) | |
| Zero Drift | ≤±0.1%T (500nm) | |
| Baseline Flatness | ± 0.005A | |
| Stability | ± 0.004A/h at 500 nm | |
| Stray Light | < 0.25% at 220 nm and 340 nm | < 0.15% at 220 nm and 340 nm |
| Scanning Capability | Yes | |
| Physical Characteristics | ||
| Standard Sample Holder | 4-position Cell Holder | |
| Analogue/Digital Output | Digital, RS232 Serial, Centronics Parallel | |
| Software | Application software | |
| Dimensions | 440(L) x 370(W) x 220(H) | |
| Weight | 17 kg | |
| Shipping Dimensions | to be updated | |
| Shipping Weight | to be updated | |
| Power Requirements | 115V/60Hz or 230V/50Hz 300VA, switchable | |
| Operating Temperature | Room temperature: 22 ± 2°C | |
| Humidity Range | ≤ 85% relative | |
| Safety Certificate | CE | |