Power: |
AC (220) (50) Hz Rechargeable nickel-hydrogen |
|
batteries which can work more than 2 hour after 15 hours fully charged.(Operate at 5ml/h) |
|
Adjustable rate range: |
|
|
|
|
|
|
|
Fixed rate: |
200ml/h ( 50ml syringe) |
|
|
|
160ml/h ( 20ml syringe) |
|
|
|
Display volume(Σ): |
0-999ml ( 0.1ml each level) |
|
|
|
Total volume(Σ): |
0-999ml ( 1ml each level) |
|
|
|
KVO: |
(0.5 - 0.3)ml/h |
|
|
|
|
|
|
Precision error range: |
2% (Injection Rate>10ml/h); |
|
|
|
3% (Injection Rate≤10ml/h) |
|
|
|
Size/Weight: |
290mm*130mm*130mm /2.1kg |
|
|
|
Rated output capacity: |
15 Va |
|
|
|
obstruction press: |
High level: 100kpa |
|
|
|
Medium level: 60kpa |
|
|
|
Low level: 40kpa |
|
|
|
Warning function: |
a.Nearly completed alarm; b.Injection completed alarm; c.Occlusion alarm; d. Reach the set volume alarm; e. Insufficient battery voltage alarm; f.Cable off alarm. |
|
Storage conditions |
a.Temperature: (-40) -55 degree; b.Relative Humidity: 20%-90%; c.Atmospheric pressure: 50kpa-106kpa |
|
|
|
|
|
|
|
|
|
|
|