Dimensions (W X D X H):550 × 460×1450㎜
Weight:Aprox. 60 Kg
Power Supply:AC220V, 50~60HZ
Connected Load:2000VA
Water Supply:Pressure Range :0~8.0BAR, Temperature Range:0~35℃
Venous Pressure Working Range:-65~100KPA(-500~+800mmhg), Accuracy±10mmhg
Arterial Pressure Working Range:-65~100kpa(-500~+800mmhg), Accuracy±10mmhg
Transmembrane Pressure Working Range:-65~100kpa(-500~+800mmhg),Accuracy±10mmhg
Blood Flow Range:0~700ml/Min
Dialysate Flow Range:0~300~500~800ml/Min (Adjustable)
Substitution Fluid Flow:0~25L/H
Dialysate Temperature Working Range:30~40℃,Accuracy±0.5℃
Heparin Pump Working Range:0~10 ml/H,Accuracy±0.1ml/H
Conductivity Working Range:10~20 ms/cm Accuracy±0.1ms/cm
Air Detector:Detecting?0.05ML/Single Bubble Detecting
Blood Leak Detector:0.5ml/l Blood Leak Detecting
Hot Rinse:more than or equal to85℃
Clean/ Disinfection:Chemical Disinfection:Sodium Hypochlorite, Peracetic Thermal Acid or Formalin
Disinfection:Hot Water, or With Citric Acid
Decalcification:Rinse,CitricAcid,Glacial Acetic Acid