Dimensions (W X D X H):550 × 460×1450㎜
Weight:Aprox. 55 Kg
Power Supply:AC220V, 50 HZ
Connected Load:Less than or equal to 2000VA
Water Supply:Pressure Range:0~8.0BAR,Temperature Range:0~35 centidegree
Venous Pressure Working Range:-65~100KPA(-500~+800mmhg), Accuracy sign10mmhg
Arterial Pressure Working Range:-65~100KPA(-500~+800mmhg), Accuracy sign 10mmhg
Transmembrane Pressure Working Range:-65~100KPA(-500~+800mmhg), Accuracysign10mmhg
Blood Flow Range:0~700ml/Min
Dialysate Flow Range:0~300~500~800ml/Min (Adjustable)
Ultrafiltration Rate:0~5000ML/H
Dialysate Temperature Working Range:30~40 centidegree ,Accuracysign0.5 centidegree
Conductivity Working Range:10~20 ms/cm Accuracy sign 0.1ms/cm
Heparin Pump Working Range:0~10 ml/H ,Accuracy sign 0.1ml/H
Air Detector:Detecting more than or equal to 0.05ML/Single Bubble Detecting
Blood Leak Detector:0.5ml/l Blood Leak Detecting
Hot Rinse:more than or equal to 85 centidegree
Clean/ Disinfection:
Chemical Disinfection :Sodium Hypochlorite, Peracetic Thermal Acid or Formalin
Disinfection:Hot Water, or With Citric Acid
Decalcification:Rinse,CitricAcid,Glacial Acetic
Acid