The Class II Biological Safety Cabinet is designed with inward air flow at a velocity to protect personnel, HEPA-filtered vertical laminar airflow for product protection, and HEPA-filtered exhausted air for environmental protection.
BSC-1100IIA2-X is designed to meet the performance criteria:
1. US Standard ANSI/NSF 49: 2002
2. European Standard EN12469: 2000
3. Chinese Standard YY0569-2005
Technical Parameters;
I: Operating Range
1. 1 Running conditions 110V— 220V / 50Hz— 60Hz
1. 2 Operating temperature range: 15-35o C
1. 3 Operating humidity range ≤ 85%
II Parameter and Function
2. 1 Filter
2. 1. 1 Inflow filter: HEPA filter,
2. 1. 2 Filter efficiency: 99. 999%(0. 1-0. 2μm)
2. 1. 3 Level of exhaust filter: HEPA filter
2. 1. 4 Exhaust filter efficiency: 99. 99%(0. 3μm)
2. 1. 5 Display of negative pressure and filter life: Bar Code
2. 1. 6 Display of accumulation working hours: VFD
2. 1. 7 Filter replacement: Front of the cabinet
2. 2 Electric Motor
2. 2. 1 Frequency conversion electric machine could make aural and visual alarm by the blockage state of the filter
2. 2. 2 Automatic adjustment of the rotating speed to keep the wind speed.
2. 3 Airflow System
2. 3. 1 Air Circulation: 70%
2. 3. 2 Exhaust Air: 30%
2. 3. 3 Inflow Velocity: 0. 65 ± 0. 15m/s
2. 3. 4 Down flow Velocity: 0. 37± 0. 015m/s
2. 4 Front Window
2. 4. 1 Sensitive Alarm: The height of the front window could adjust at will, when overtop the safety height, it will alarm
2. 4. 2 Safety Glass: Toughened glass, explosive-Proof and prevent UV
2. 4. 3 Thickness of front panel is more than 5mm
2. 5 Workroom Interior
2. 5. 1 Material: Table-board is made of excellent stainless steel, two flanks and back side are punched into a smooth whole body, more easy to cleaning.
2. 5. 2 Power Socket: 2 internal waterproof socket
2. 5. 3 Lighting > 800lux
2. 5. 4 UV Lamp: Sterilizing by UV lamp, the UV light couldn't turn on when the fluorescent lamp turning on or the front window open
2. 5. 5 Air Grill: Air will be drawn into the cabinet through the air grill.
2. 5. 6 Water and gas tap
2. 6 Size
2. 6. 1 internal size(w *d *h): 970x600x600mm
2. 6. 2 external size(w *d *h): 1100x800x2200mm
2. 7 Noise: < 65 dB
2. 8 Power: 0. 45Kw
2. 9 Weight: 180Kg
2. 10 Vibration: The net displacement is no more than 3μm when the frequency is from 10Hz to 10KHz
2. 11 Stability: The core part of work surface can stand more than 23kg pressure without permanent distortion.
2. 12 Biosafety
2. 12. 1 Personnel safety: Hammerblow samples ≤10CFU/time
Slit samples ≤5CUF/time
2. 12. 2 Product safety: The microorganism germ falls ≤5CUF/time
2. 12. 3 Cross infection safety: The microorganism germ falls ≤2CUF/time
III Configure
3. 1 Main Unit
3. 2 Inflow ULPA
3. 3 Exhaust HEPA
3. 4 UV lamp
3. 5 Light
3. 6 Socket
IV Accessories
4. 1 Water Valve
4. 2 Gas Valve
4. 3 Vacuum Valve
4. 4 Removable Stand
4. 5 Remote control