Wavelength Range
|
190nm~900nm
|
|||
Wavelength Accuracy
|
±0.15nm
|
|||
Wavelength Repeatability
|
≤0.05nm
|
|||
Spectral Bandwidth
|
0.1nm, 0.2 nm, 0.5 nm, 1.0 nm, 2.0 nm
|
|||
Accuracy of spectral band shift
|
±0.3nm
|
|||
Resolution
|
<30%
|
|||
Edge Noise
|
<0.02Abs
|
|||
Edge Energy
|
<2%
|
|||
Baseline Stability
|
Static Baseline
|
Drift
|
±0.002Abs
|
|
Noise
|
<0.002Abs
|
|||
Ignition Baseline
|
Drift
|
±0.006Abs
|
||
Noise
|
≤0.006Abs
|
|||
Character Concentration
|
Cu<0.02μg/mL/1%(Air-Acetylene Analysis, Air -Natural Gas or Liquefied Petroleum Gas Analysis)
|
|||
Ba<0.15μg/mL/1%(Nitrous Oxide-Acetylene Analysis)
|
||||
Character Value
|
Cu<2.0×10-11 g
|
|||
Cd<3.0×10-13 g
|
||||
Detection Limit
|
Flame
|
Cu<0.004μg/mL(Air-Acetylene Analysis, Air -Natural Gas or Liquefied Petroleum Gas Analysis)
|
||
Grphite Furnace
|
Cd<4.0×10-13 g
|
|||
Repeatability
|
Flame
|
Cu<0.6%(Air-Acetylene Analysis)
|
||
Cu<1.0%(Air –Natural Gas or Liquefied Petroleum Gas Analysis)
|
||||
Ba<1.0%(Nitrous Oxide-Acetylene Analysis)
|
||||
Grphite Furnace
|
Cu<2%, Cd<2%
|
|||
Background Correction
|
>×55(Deuterium Lamp, Self-Reversal)
|
|||
Spray volume of sample solution and Atomization efficiency
|
Spray Volume>5mL/min;
|
|||
Atomization Efficiency≥8%
|
||||
Аbsorbancy Accuracy
|
±1%
|
|||
Sensitivity
|
Flame
|
≥0.250Abs
|
||
Grphite Furnace
|
≥0.08Abs
|
|||
Lamp Turret
|
Eight Hollow Cathode Lamps
|
|||
Analytical Method
|
Flame(Air-Acetylene Analysis, Nitrous Oxide-Acetrlene, Air -Natural Gas or Liquefied Petroleum Gas Analysis), Grphite Furnace
|
|||
Heating program(Grphite Furnace)
|
Heating Profile with up to 20 Steps
|
|||
Communication Type
|
USB,RS232
|
|||
Safety Function
|
Burner Identification,Flame Sensor, Gas Leak Sensor, Low Gsa Pressure Sensor, Power Loss Protection, Gas Loss Flane Protection.
|
|||
Power Supply
|
Compatible 110V and 220V(Switchable)
|