Product Main

Specifications

APPLICATION INSTRUCTION FOR DIGITAL INDICATOR
IMPORTANT
1. The response speed: ≤0.6m/s.
2. Avoid shocks, impact, dropping etc., otherwise it will damage the appearance and precision.
3. Keep indicator face clean, prevent liquid from getting into inside.
4. Do not remove output connector lid without using output interface. Output end should never be connected with metal. Doing so could damage your indicator.
5. Do not connect any part of digital indicator with voltage; otherwise it will damage the circuit.
6. Must take out the battery if not in use for a long time.

SPECIFICATIONS

Resolution: digital indicator: 0.01mm/0.0005″,
digital micro-indicator: 0.001/0.00005″.
Measuring Range: digital indicator: 0~12.7mm/0.5″,0~25.4mm/1″
digital micro-indicator: 0~5mm/0.2″, 0~12.7mm/0.5″,
0~25.4mm/1″
Power: One sliver oxide battery (SR44) with work voltage 1.5v.
Operating temperature: 0℃~40℃
Storage temperature: -20℃~70℃
Relative humidity:≤80%

NOMENCLATURE

  
1. Contact point
2. Spindle
3.Mounting diameter
4. On/off switch
5. Zero-setting
6. In/mm switch
7. LCD display
8. Data output interface
9. Battery compartment
10. Anti-dust-cap
11. The even back cover (or with back lug)
FUNCTION
1. Zero setting at any position;
2. "ON/OFF” button;
3. Metric/Inch system interchange at any position;
4. With two type data output interface: normal or mini-USB interface.
Usage:
1. Clean the contact point and spindle with a soft cloth before using.
2. Press the OFF/ON switch. The 0.00mm or 0.000mm display
appears, and the indicator is now ready for measurement.

The method of replacing battery

Remove battery compartment lid and replace one SR44 battery with its positive (+) side facing upward.

TROUBLESHOOTING

Failure
Possible Situation
Solutions
All digits appear simultaneously and change rapidly
Battery power low
Remove battery compartment lid and replace battery
The display does not work (only when changing battery)
Circuit accidental pause caused by poor voltage when fixing battery
Remove battery and replace it after 30 seconds
Reading error for full length is ≤0.1mm
Dirt has damaged/covered sensor
Remove cover and slider assembly. Apply a small layer of petrol to avoid dirt build-up
Nothing appears on the LCD
Battery has poor contact. Battery voltage is under 1.1v
Keep good contact for battery or replace the battery