CHARACTERISTICS:
1> 2 control ways: Can be used in either drop counting control way or volumetric control way ml/h or drops/min
2> High speed: 1ml/h~1000ml/h or 1 drop/min ~250 drop/min.
3> 3 levels of occlusion value: High, middle and low.
4> Hard based free-flow protection: Avoid any free flow or leakage, prevent any danger to patients.
5> Drops detector: Can detect the drops infused, automatically detect the IV line type and ensure the accuracy for any IV line.
6> Applicable IV Sets: Any standard IV sets.
7> Rechargeable internal battery.
8> Visible and audible alarm: When any malfunction, error, air-on-line, occlusion, error flow rate and so on.
9> KVO function: After infusion is complete, the pump automatically switches to the smallest flow to prevent creating the thrombus.
10> Bolus function: When during infusion, press a key to infuse at the rate of 400ml/h.
SPECIFICATIONS:
1> Accuracy: +/-3% ~5%
2> Delivery Rate Range: 1~250 drop/min(1drop/min step) or 1ml/h~1000ml/h
3> Volume Limit Range: 0~9999ml (1 ml step)
4> Bolus rate: 500ml/h
5> KVO Rate: 1ml/h~5ml/h
6> Alarms: Air-in-line, Infusion complete(a volume limit has been set), Empty container, Occlusion, Door open, Over time, No drop sensor, Operation error, Low battery and other device failures.
7> Power supply: AC100~ 240V+/-10%, 50/60Hz; DC 12V
8> Power Consumption: ≤ 28VA
9> Shock Protection: Class I, Type BF
10> Dimensions: 135mm*120mm*245mm
11> Net Weight: < 2.4 kg
12> Operating Condition: Ambient temperature: 5 to 40C Relative humidity: 20 to 90%
Storage Condition: Ambient temperature: -10 to 55C Relative humidity: 10 to 90%
13> Accessories: AC power line, Operation manual