Product Main

Specifications

Technical Data:      
Tube Voltage/Current: 65kv/1.0mA
Power Supply: 220V±5%      50HZ
Focal Spot to Skin Distance: 20cm
Irradiation Diameter: 6cm
Focus: 0.5mm~0.8mm
Total Filter Amount: 1.5mm aluminum equivalent
Exposure Time: 0.01s~2.50s, 0.12s for warm up   Rocker Horizontal Angle: ±600
Rocker Travel Up and Down: ≥400mm
Current of Whole Device: ≤1A
Tube Head Movement: ≥300mm
Tube Head Rotate Along Vertical Shaft: 3600
Power    250VA
Weight of Main Unit    25 kg
Gross Weight       30kg