Product Main

Specifications

Westingarea supplied high quality Transport Ventilator has the excellent performance displayed as

Classical emergency transport ventilator
Pneumatically driven pneumatically controlled, not rely on electric power
With A/C mode and pressure trigger
With pressure alarm
Oxygen concentration adjustable
Adult/pediatric parameter in different color, easy to operate
Simple hose connecting, time saving
Compact and portable
Can be used at emergency department, emergency station, county hospital, clinic and for patients inter-departmental transfer
 
Certification :  CE

Parameter  MV 3~20L/min(used in 60%), 3~14L/min(used in 100%); Respiratory rate 1060bpm; IE ratio 12; FiO2 60%, 100%; Trigger sensitivity Pressure, -3cmH2O
 
Monitoring  Pressure values Paw; Alarm Paw high, Paw low, Apnea, O2 supply down
 
Technical data  Screen LCD; Gas supply O2, 0.25~0.6MPa; Maximum security pressure 8kPa; Compliance ≤4ml/100Pa
 
Noise   65dB(A)
 
Communication interfaces   377*110*142mm
 
Weight   2.4kg
 
Environment requirement  Temperature -1850(Operation), -2055(Storage);Relativehumidity 5%95%non-condensing(Operation), ≤80%, non-condensing(Storage)
 
Atmospheric Pressure   0110kPa(Operation), 50110kPa(Storage)