Characters of our Clinical infant incubators
Microprocessor based servo controlled temperature system
Air temperature,skin temperature,humidity,timer and heat power separately displayed by LED
Alarms functions : Power failure; over temperature; temperature deviation; temperature sensor failure;fan failure;
>37 temperature setting
A keypad lock helps prevent inadvertent changes the setting
Temperature deviation and over temperature could be amended in front panel directly
Heater is electrically cut-out if air temperature inside incubator exceed 40
The infant bed inclination angle is adjustable
Humidity is adjustable
Trough for pulling and pushing ,which is convenient to add the water and wash
Fan with low noise and long life-span
six operating windows,two side operating windows can be swiveled
Integral timer to record therapy time
Skin temperature sensor failure protect function avoid over temperature
cabinet with three drawers, silence castors
A second thermal cut-off function for more safety
Large size and double wall hood
Standard configuration
Main body(including the transparent hood,control system,infant bed,water trough),skin temperature sensor,I.V.pole, mattress, cabinet, castors
Optional parts: Phototherapy up side unit; phototherapy from up and down sides;RS232(485) connector ;Tray/pole
Technical data:
Power supply : 110/220V 60/50Hz
Power input : ≤420VA
Operating conditions:
Environment temperature: 20~30
Environment relative humidity: 30%~70%RH
Environment air velocity of flow: <0.3m/s
Air Temperature control range 25~38
Skin temperature control range:32~38.5
Skin temperature sensor accuracy: +0.3
Temperature fluctuation: +0.5
Uniformity of mattress temperature: ≤0.8
Warm-up time: ≤30min
Internal noise level: ≤48dB
Humidity display range:10%-95%RH
Humidity accuracy: ≤10%RH
Infant bed tilt angle : +10°
Mattress size: 67cm(L)*54cm(W)
Trough capacity: 1200ml
Air filter:0.5um
Package:
Each unit is packed in two wooden crate
Size (control unit and cabinet):123*69*106CM N.W.:85KG G.W.:125KG
Hood: 97*62*68CM N.W.:15KG G.W.:40KG
Transport and storage
Environment Temp:-10--+70
Environment Relative Humidity: ≤80%RH
Atmospheric Pressure: 500-1060Hpa