Dynamic strain indicator:
Duct No.:can be united any ducts(for example: 4, 6,8 duct).
Electric bridge resistance: 60~1000Ω
Electric bridge voltage: 2、4、6、8 (controlled by switch)
Balance range: Electric bridge resistance ±1%(±5000μ)
Sensitivity: 120mV/10μ(BV=4V)
Output: output voltage: ±10Vp±5mV; output current flow: ±10Ap±30mA
Output I sensitivity adjustment: 1~100%
Sensitivity coefficient: 2.00
Output impedence: 1Ω
Nonlinear: ±0.1%F.S
Calibration strain: 10~9990μ(three position dialing switch)
Calibration precision: ±(0.5%±0.5μ)
Sensitivity adjustment: 0、1、2、5、10、20×100μ 6step ,slight adjustment 1~3times
frequency response characteristic:
DC linking 0~50kHz(±0.5dB)
0~150kHz(-3dB±0.5 dB)
AC linking 5Hz~50kHz(±0.5dB)
1.5Hz~150kHz(-3dB±0.5dB)
Low pass filter: 100、300、1k、3k、10k、F(150k)Hz
Rate of decay: 12dB/oct
Transition location: -3±1dB
Signal to noise ratio:
100μrange ≤46dB
Other range ≤52dB
Zero shifting:
±0.1μ/degree
±0.5μ/24hour
Working condition:
Working temperature -10degree~+50degree
Working temperature ≤85%RH(no freezing)
G-load: >±10Vp
Power:
AC 220V±10% 8VA/duct
DC 10.5~15V 0.5A/duct