Product Main

Specifications

*Single channel with measurement report
*Build in rechargeable Ni-MH battery
*Chinese/English language
*Simultaneous 12 leads
*320X240 graphic LCD, 3.8 inch screen
*Paper size: 50mmX20/30m
*Built in RS232 interface (USB is optional)
*8 Patients save/copy/communication


Specification

Lead: standard 12 leads
lead acquisition: synchronously 12 leads
Data acquisition: 18bit/1000Hz
Record mode: Manual mode; Auto 1; Auto 2; Auto 3
Filter: AC, EMG,Drift Filter
CMRR :100dB(with AC filter)
Input circuit: Floating; Protection circuit against Defibrillator effect
Input circuit current :≤0.1μA
Input Impedance: ≥2.5MΩ
Patient current leakage: 3.2s
Frequency response: 0.05 ~150Hz
Noisy level :≤15μVp-p
Sensitivity valve: ≤20μV
Inter-Channel Interference :≤0.5mm
Paper speed :25,50mm/s(+/-3%)
Sensitivity :AUTO,2.5,5,10,20mm/mV
Recording mode :Thermal Array printing
Paper specification :50mmx20m,roll
LCD display :320x240 graphic LCD,display status&waveform.
Safety classification: IEC60601-1 class I, type CF
Power supply :AC :110V/220V/85~265V,50/60Hz +/-1Hz;30VA
DC :12V(1500mAh) rechargeable Ni-MH battery
Fuse: AC extend T500mA/125V     OD5x20mm
AC extend T250mA/250V     OD5x20mm
AC extend T2.0A/250V         OD5x20mm