The following details is for standard of AC/DC 150KV, parameters can be made according to user's requirements.
FCR-150 high Voltage Divider is a universal high Voltage measuring instruments which can
be used to test Power frequency AC high voltage and DC high voltage for electric power
systems, electrical appliance, electronic equipment manufacturing sector. It includes
high voltage testing part & low-voltage measuring instruments. The low-voltage & high voltage part can work separately, so it is of safety and reliability. It is the
upgrading products of static high voltage meter, packed with aluminum alloy case for
portable.
be used to test Power frequency AC high voltage and DC high voltage for electric power
systems, electrical appliance, electronic equipment manufacturing sector. It includes
high voltage testing part & low-voltage measuring instruments. The low-voltage & high voltage part can work separately, so it is of safety and reliability. It is the
upgrading products of static high voltage meter, packed with aluminum alloy case for
portable.
Technical Parameters
1.Measuring range: AC 150KV; DC 150KV
2.Testing Accuracy: AC 1%, DC 0.5%
3.Impedance: 1800MΩ
4.Divider rato: 1:1000
5.Condition: Temperature 0~40℃; Humidity:<85%
Methods of use
Model
|
Testing
Range
(kv)
|
Impedance
(MΩ)
|
Size (mm)
|
Weight
(kg)
|
|
Divider
|
Alloy Case
|
||||
FRC-50
|
AC-50 DC-70
|
1000
|
Ф76×450
|
220×220×670
|
8
|
FRC-100
|
AC-100 DC-140
|
1500
|
Ф76×700
|
220×220×750
|
10
|
FRC-150
|
AC-150 DC-200
|
1800
|
Ф76×1000
|
220×220×1100
|
15
|
FRC-200
|
AC-200 DC-250
|
4000
|
Ф76×1300
|
220×220×1400
|
18
|
FRC-300
|
AC-300 DC-400
|
6000
|
Ф76×1700
|
220×220×1800
|
25
|
Methods of use
1. Please connect with the Divider and DUT(Device under Testing)
2. The Ground wire will be connected with the Socket of Divider Base;
3. The grounding clip must be absolutely connected to ground.
4. The high voltage wire connect with the Output showed on base, then fasten High voltage
wire with the screw bolt.
wire with the screw bolt.
5. Connecting the Multimeter with the Multimeter wire;
6. Switching the Multimeter conversion button to AC200V or DC200V( “V” to be read as "kilovolt" as numerical reading).
7. Start to connect High Voltage, the digital instrument will directly show the testing data.
(Note: The staffs are not allowed to close the High Voltage Arm to avoid danger during testing).
(Note: The staffs are not allowed to close the High Voltage Arm to avoid danger during testing).
8. After completing testing, firstly cut off supply power of HV equipment and the staffs can allow
to enter the testing field until the Multimeter show the figure "0".
to enter the testing field until the Multimeter show the figure "0".