Product Main

Specifications

Composed by optical fiber  connector protective cap
 specification
(μm)
optical fiber dia.
 
Disposable sterile medical
 Laser optical fiber 
YYGX200,YYGX2001,YYGX2002,YYGX20012
200μm±10%
YYGX400,YYGX4001,YYGX4002,YYGX40012
400μm±10%
YYGX600,YYGX6001,YYGX6002,YYGX60012
600μm±10%
 
Non-sterile
medical
 Laser optical fiber 
YYGX200F,YYGX2001F,YYGX2002F,YYGX20012F
200μm±10%
YYGX400F,YYGX4001F,YYGX4002F,YYGX40012F
400μm±10%
YYGX600F,YYGX6001F,YYGX6002F,YYGX60012F
600μm±10%
 
1.Appearance: fiber surface should be clean,seamless and burrfree
2.Transmission efficiency: should be greater than 70%
3.folding  strength
(um)
Fiber dia.
200
400
600
(mm)
Folding dia.
≥25
≥35
≥45
biological  property of Medical laser optical fiber  
4.axenic : (One-off axenic medical Laser optical fiber)
5.Cytotoxicity:  not greater than level I
6.intracutaneous stimulation:  not greater than slight stimulation
7.delayed type hypersensitivity: should be no this  hypersensitivity
8.Hemolysis: not greater than 5.0%
9.systemic toxicity test:  should be no this  toxicity
 
Chemical property of Medical laser optical fiber
10.Reducing substances :potassium permanganate consuming volume difference between test liquid and comparing liquid Should not be greater than 2ml
 11.pH value:  pH value difference  between test liquid and comparing liquid Should not be greater than 1
 12.heavy metal concentration: total heavy metal concentration in test liquid  Should not be greater than 5μg/ml,cadmium   concentration not be greater than 0.1μg/ml
 13.Epoxy Ethane Residue: should not be greater than 10μg/g in optical fiber Sterilized by Epoxy Ethane (only apply to  Disposable sterile medical Laser optical fiber)