| Dimension:288mmX204mmX60mm |
| Packing:380mmX300mmX210mm |
| Weight:4.3kg(gross); 3.27kg(net) |
Single channel with measurement report
Build in rechargeable Ni-MH battery
Chinese/English language
Simultaneous 12 leads
20X2 character LCD
Paper size: 50mmX20/30m
Built in RS232 interface (USB is optional)
Specification
|
Items
|
specification
|
|
Lead
|
standard 12 leads
|
|
lead acquisition
|
synchronously 12 leads
|
|
Data acquisition
|
18bit/1000Hz
|
|
Record mode
|
Manual mode; Auto 1; Auto 2; Auto 3
|
|
Filter
|
AC, EMG, Drift Filter
|
|
CMRR
|
100dB(with AC filter)
|
|
Input circuit
|
Floating; Protection circuit against Defibrillator effect
|
|
Input circuit current
|
≤0.1μA
|
|
Input Impedance
|
≥2.5MΩ
|
|
Patient current leakage
|
<10μA
|
|
Calibrating voltage
|
1mV±2%
|
|
Voltage Tolerance
|
±500mV
|
|
Time constant
|
>3.2s
|
|
Frequency response
|
0.05 ~150Hz
|
|
Noisy level
|
≤15μVp-p
|
|
Sensitivity valve
|
≤20μV
|
|
Inter-Channel Interference
|
≤0.5mm
|
|
Paper speed
|
25, 50mm/s(±3%)
|
|
Sensitivity
|
AUTO, 2.5, 5, 10, 20mm/mV
|
|
Recording mode
|
Thermal Array printing
|
|
Paper specification
|
50mmx20m, roll
|
|
LCD display
|
20x2 character LCD, display status.
|
|
Safety classification
|
IEC60601-1 class I, type CF
|
|
Power supply
|
AC : 110V/220V/85~265V, 50/60Hz±1Hz; 30VA
|
|
DC : 12V(1500mAh) rechargeable Ni-MH battery
|
|
|
Fuse
|
AC extend T500mA/125V ¢5x20mm
AC extend T250mA/250V ¢5x20mm
AC extend T2.0A/250V ¢5x20mm
|