Product Main

Specifications

1.             IV Set Management and Flow Rate Calibration
IV sets can be added, stored and deleted in memory, and easily used after calibration.
2.             Optional Infuse Rate Unit
d/min, mL/h and volume/time
3.             Infusion Mode
Drip-rate and volumetric
4.             Pumping Mechanism
Peristaltic finger pump
5.             IV Sets
Standard IV sets of 10, 15, 20 and 60d/mL, with diameter 3.4 ~ 4.5mm.
6.             Flow Rate Range
1.0 ~ 1200 mL/h
7.             Drip Rate Range
10 d/mL IV set: 1.0 ~ 200 d/min
15 d/mL IV set: 1.0 ~ 300 d/min
20 d/mL IV set: 1.0 ~ 400 d/min
60 d/mL IV set: 1.0 ~ 150 d/min
8.             Increment
From 0.1 mL/h to 99.9 mL/h, 0.1 mL increment; ≥100 mL/h, 1 mL increment.
9.             Flow Rate Accuracy
≤ ± 5% (after calibration)
10.        Drip Rate Accuracy
≤ ± 1.5%
11.        Volume Limit
1 ~ 9999 mL
12.        Volume Infused Display
0 ~ 9999 mL & >9999 mL
13.        Purge Rate
600 mL/h
14.        KVO Rate
1 mL/h
15.        Bolus Mode
Available for Bolus mode, flow rate adjustable.
16.        Occlusion Pressure
Pressure detection device with 8 sensitivities setting, 40 ~ 140KPa.
17.        Air-in-line Detection
Air bubble detected by ultrasonic and the device with 8 sensitivities setting, 50 ~ 1000μL.
18.        Alarm Volume
8 levels setting
19.        Anti-reverse Function
Anti-reverse function on motor to prevent upstream
20.        Self-check
Self-check when start up and during infusion
21.        Audio and Visual Alarms
Air-in-line, Door open, Occlusion, Infusion finished, Pause overtime, Leakage, Empty, Low battery, Battery damaged, etc.
22.        Display
Blue LCD with high brightness
23.        Power Source
AC: 100V ~ 240V, 50Hz/60Hz, ≤ 25VA
24.        Battery    
Rechargeable Ni-MH battery: DC12V, 2000mAh, life: ≥ 4 hours at 25mL/h
25.        Classification
Class Ⅱ, CF, IPX4
26.        Dimensions and N.W.
150mm x 150mm x 235mm; 2.3kg