Function and Features
1). Flow rate range: from 1-600ml/h.
2). Unique door-free structure: making it easier for user to handle and
avoiding problems of pump stuck and fluid leakage
3). Big LCD screen: making it easier and more convenient to operate
4). 4-level air alarm control: a great solution to avoid frequent air alarms
5). Unique bedside infusion supervision function: making patients and their relatives more assured while reducing nurses' workload.
6). Open system tor IV set: accept all brands of standard IV sets with only a simple calibration to ensure accuracy. Can keep values for 3 brands/types of iv sets
7). Audio-visual alarms: alarms given in case of air in line or occlusion with machine stopping work automatically.
8). Ultrasonic sensor: accurate air alarm given when detecting excessive air pressure in line
9). Double CPU: making infusion process safer and more reliable.
10). Alarm signals are given in case of malfunction or operation error and the machine stops operation automatically.
11). can add feeding function in the same pump if required by customer
|
Dimension
|
120*140*190mm (length*width*height)
(outer shell: ABS plastic) |
|
Power supply
|
AC100-240V,50/60 Hz,25VA
|
|
|
Battery
|
Rechargeable lithium polymer battery, 7.4V, 1600mA
|
||||
|
Weight
|
<2.0kg
|
|
|||
|
Applicable infusion set
|
Accept all brands of IV sets of national standard
|
||||
|
Fuse
|
220V 2A*2, 12V 2A*2
|
||||
|
Accuracy
|
±5%
|
||||
|
Electrical safety
|
Comply with the stipulation of GB9706.1-1995, EN60601-1 and UL60601-1
|
||||
|
Flow rate range
|
1-600ml/h
|
||||
|
Infusion increment
|
1ml/h
|
Batteryrecharge
|
When the pump is connected to the AC power supply, turn on the power and the battery will automatically recharge (about 8-14 hours to recharge fully).
|
|
|
|
Preset volume limit
|
1-9999ml
|
|
|||
|
Displayed information
|
Rate, volume limit, accumulated volume, power indicator light, bed No., AIR, OCCL., OVER.
|
Maximum power consumption
|
25VA, can run for 5h after being recharged
|
|
|
|
Operating conditions
|
Environment temperature +5ºC ~ +40ºC,atmosphere pressure 86~106kPa, relative humidity20%~90%
|
||||
|
Bolus rate
|
600ml/h
|
|
|||
|
KVO flow rate
|
1ml/h,keep vein open(KV))rate
|
Storage and transport conditions
|
Environment temperature -15ºC ~ +50ºC , atmosphere pressure 86~106kPa, relative humidity20%~90%.
|
|
|
|
Alarm function
|
Over, OCCL.,AIR, low battery, control abnormal, no AC power supply, installation error.
|
Electromagnetic compatibility
|
Comply with EN60601-1-2
|
||
|
Work mode
|
Continuous work mode
|
Applied standards
|
Comply with standard of EN ISO 14971:2000,IEC 60601-1:1995,EN IEC 60601-1-2:2001,IEC 60601-2-24,EN ISO 14155-1:2003,EN 980:2003,EN 1041:1998,MDD 93/42EEC
|