Product Main

Specifications

Function and Features  
1). Flow rate range: 0.1-2000 ml/h
2). 3 modes: Rate mode, Time mode, Weight mode
3). Unique door-free structure: making it easier for user to handle and avoid problems of pump stuck and fluid leakage
4). Big LCD screen: making it easier and more convenient to operate
5). Unique bedside infusion supervision function: making patients and their relatives more assured while reducing nurses' workload.  
6). Open system tor IV set: accept all brands of standard IV sets with only a simple calibration to ensure accuracy. Can keep values for 3 brands/types of iv sets
7). Audio-visual alarms: alarms given in case of air in line or occlusion with machine
 stopping work automatically.  
8).  KVO: selectable, 1.0 ml/h-5.0 ml/h
9). Infrared sensor: accurate air alarm given when detecting excessive air pressure in line.      
10). Double CPU: making infusion process safer and more reliable.
11). Alarm signals are given in case of malfunction or operation error and the machine stops operation automatically.
 
Dimension
120×140×190 mm
 
Accumulated infusion volume
1.0~9999.0 ml
Maximum weight
< 3.0 kg
Infusion accuracy
±5%
Displayed information
Infusion rate, preset volume, accumulated infusion volume, battery capacity, AC power indicator, infusion completion, occlusion, air bubbles, bed No., pressure value, flow rate.
Alarm function
Over,occlusion,air bubbles,low battery,installation error of infusion pipe,machine malfunction,empty
Fuse
T 2AL 250V~
KVO rate
1.0 ml/h-5.0 ml/h
Maximum power consumption
25VA, running more than 4 hours after being charged
Selectable modes
Rate mode, Time mode, Weight mode
Infusion rate range
0.1-2000 ml/h
Rate mode
1.0~2000.0 ml/h
Bolus
Bolus: 1000.0 ml/h when Rate is ≤1000.0 ml/h
Time mode
Volume limit: 0.1~999.9ml  Time: 1~2000 minutes
Operation conditions
Environment temperature 5ºC ~ 40ºC, air pressure 86kPa~106kPa, relative humidity ≤80%
Body Weight mode
Drug limit: 1.0~999.9 mg, Volume limit: 0.1~999.9 ml. Weight: 0.1~300.0 kg, Dose: 0.1~9999.9
Selectable unit: mg/kg/h, ug/kg/min
Electrical Safety
Comply with EN60601-1 for medical equipment and EN60601-2-24
Battery
Rechargeable lithium polymer battery, 7.4V, 1600 mA
Applicable infusion sets
All standard IV sets
Storage
Environment temperature-20ºC~55ºC, air pressure 50~106kPa, relative humidity ≤95%