Function and Features
1). 3 kinds of modes: Rate Mode, Time Mode, Body Weight Mode.
2). Double CPU: making injection process safer and more reliable.
3). Accurate & sensitive pressure sensor: detecting occlusion pressure accurately.
4). Injection time, dose, solution: easy to be checked during injection process.
5). Precision step motor: no pulsating wave even during low-rate injection.
6). KVO & Bolus functions: rate selectable according to actual clinical requirement.
7). Excellent injection operation: accuracy ensured even at small-volume injection.
8). Unique injection supervision system: ensures safe injection and reduces the nurse’s workload.
9). Open system for syringes: accept all brands of syringes of national standard from 5ml, 10ml, 20ml, 30ml to 50ml, which can be self-diagnosed and displayed on LCD display.
10). Alarms: audio-visual alarms given in case of malfunction or operation error with machine stopping at the same time automatically.
11). Preset volume: automatically saved in case of unpredictable switch-off, thus ensuring accuracy of injection volume.
12). Drug name library: As many as 200 drug names can be stored in the machine and can be checked on LCD display during operation.
|
Dimension
|
288x130x122mm (outer shell: ABS plastic)
|
|
Power supply
|
50/60Hz AC100V~240V
|
|
|
Weight
|
<3.0kg
|
Battery
|
Rechargeable lithium polymer battery, 7.4V, 1600mA
|
||
|
Selectable modes
|
Rate mode, Time mode, Body Weight mode
|
||||
|
Fuse
|
220V2Ax2, 12V2Ax2
|
||||
|
Applicable syringes
|
Accept all brands of standard syringes:
5ml, 10ml, 20ml, 30ml, 50ml |
Electrical safety
|
Comply with the stipulation of GB9706.1-1995, EN60601-1 and UL60601-1
|
|
|
|
Accumulated injection volume
|
0~9999.9ml
|
Maximum power consumption
|
25VA, can run for 5h after being recharged
|
||
|
Injection accuracy
|
±2%
|
||||
|
Feasible syringe
rate range |
5ml: 0.1~100ml/h 10ml: 0.1~200ml/h
20ml: 0.1~400ml/h 30ml: 0.1~600ml/h 50ml: 0.1~1500ml/ h |
Batteryrecharge
|
When the pump is connected to the AC power supply, turn on the power and the battery will automatically recharge (about 8-14 hours to recharge fully).
|
||
|
Displayed Information
|
rate, volume limit, syringe size, pressure value, battery capacity, unit of medical dosage, bed No. AC power indicator, accumulated volume, drug name
|
|
|||
|
Alarm function
|
OVER, NEAR, Occlusion, Low battery, Syringe disengaged, Syringe abnormal, Control abnormal, Internal malfunction
|
||||
|
Electromagnetic compatibility
|
Comply with the stipulation of IEC60601-1-2 (1993 edition) EN60601-2-24
|
||||
|
Bolus rate
|
5ml: 100ml/h, 10ml: 200ml/h,
20ml: 400ml/h, 30ml: 400~600ml/h (adjustable) 50ml: 400~1500ml/h (adjustable) |
||||
|
Operating conditions
|
Environment temperature 5ºC~45ºC, air pressure 70~106kpa, relative humidity 30%~90%
|
|
|||
|
KVO rate
|
0.1ml/h~5.0ml/h(selectable)
|
Storage
|
Environment temperature -15ºC~50ºC, air pressure 70~106kPa, relative humidity 30%~90%
|
|
|
|
Occlusion value
|
400mmHg~900mmHg
|
Water proof
|
IPX1
|
|
|
|
Time mode
|
Vol limit: 0.1~999.9ml Time: 0~2000 min
|
Classification
|
Class II, type BF
|
|
|
|
Drug name library
|
200 drug names can be stored in the machine.
|
Certification
|
CE0197
|
|
|
|
Body Weight Mode
|
Drug limit: 0.1~9999.9(mg/ug), Solution limit: 0.1~999.9ml, Weight: 0.1~300.0kg, Dose: 0.1~9999.9 Selectable unit: mg/kg/h, ug/kg/min
|
Applied standards
|
Accord with the stipulation of YZB/Yue0482-2004
|