Function and Features
1). Rate Mode .
2). Double CPU: making injection process safer and more reliable.
3). Accurate & sensitive pressure sensor: detecting occlusion pressure accurately.
4). Precision step motor: no pulsating wave even during low-rate injection.
5). Bolus & KVO functions
6). Excellent injection operation: accuracy ensured even at small-volume injection.
7). Unique injection supervision system: ensures safe injection and reduces the nurse’s workload.
8). Open system for syringes: accept all brands of syringes of national standard from 10ml, 20ml, to 50ml, which can be self-diagnosed and displayed on LCD display.
9). Alarms: audio-visual alarms given in case of malfunction or operation error with machine stopping at the same time automatically.
10). Preset volume: automatically saved in case of unpredictable switch-off, thus ensuring accuracy of injection volume.
|
Dimension
|
380*190*180mm (outer shell: ABS plastic)
|
|
Electrical safety
|
Comply with the stipulation of IEC60601-1:1996
|
|
|
Weight
|
<3.0kg
|
|
|||
|
Applicable syringes
|
Accept all brands of standard syringes:
10ml, 20ml, 50ml |
Maximum power consumption
|
25W, can run for 4h after being recharged
|
|
|
|
Accumulated injection volume
|
0.1-999.9ml
|
||||
|
Battery recharge
|
When the pump is connected to the AC power supply, turn on the power and the battery will automatically recharge (about 8-14 hours to recharge fully).
|
||||
|
Injection accuracy
|
±3%
|
|
|||
|
Feasible syringe
rate range |
10ml:0.1ml/h-100ml/h;
20ml:0,1ml/h-200ml/h; 50ml:0.1ml/h-500ml/h |
Electromagnetic compatibility
|
Comply with the stipulation of IEC60601-1-2 :2006
|
|
|
|
Displayed Information
|
Rate, Accumulated injection volume, Syringe size, Battery capacity, Bed No., AC Power indicator, Battery charge indicato
|
Operating conditions
|
Environment temperature +5ºC~+40ºC,atmosphere pressure 86~106kPa, relative humidity ≤80%
|
|
|
|
Storage
|
Environment temperature -40ºC~+55ºC, atmosphere pressure 50~106kPa, relative humidity≤95%.
|
||||
|
Alarm function
|
Over, NEAR, Occlusion, Low battery, Syringe disengaged, No AC power supply, Serious low battery
|
||||
|
Classification
|
Class II, type BF
|
||||
|
Bolus rate
|
The fastest rate of the size of syringe in use
|
||||
|
Certification
|
CE0197
|
||||
|
Power supply
|
50/60Hz AC100V~240V
|
||||
|
Applied standards
|
IEC 60601-1:2005
IEC60601-2-24:1998 IEC60601-1-2:2006 ISO780-1997 ISO15223:2000 ISO7886-1:1993 ISO/IEC:Guide No.37 <Consumer Goods Usage Instructions> |
||||
|
Battery
|
Rechargeable lithium polymer battery,
7.4V, 1600mA |
|
|||
|
Fuse
|
T2AL250V
|
|