Product Main

Specifications

Function and Features 
1). Rate Mode .
2). Double CPU: making injection process safer and more reliable.
3). Accurate & sensitive pressure sensor: detecting occlusion pressure accurately.   
4). Precision step motor: no pulsating wave even during low-rate injection.
5). Bolus & KVO functions
6). Excellent injection operation: accuracy ensured even at small-volume injection.
7). Unique injection supervision system: ensures safe injection and reduces the nurse’s workload.
8). Open system for syringes: accept all brands of syringes of national standard from 10ml, 20ml, to 50ml, which can be self-diagnosed and displayed on LCD display.
9). Alarms: audio-visual alarms given in case of malfunction or operation error with machine stopping at the same time automatically.
10). Preset volume: automatically saved in case of unpredictable switch-off, thus ensuring accuracy of injection volume.
Dimension
380*190*180mm (outer shell: ABS plastic)
 
Electrical safety
Comply with the stipulation of IEC60601-1:1996
 
Weight
<3.0kg
 
Applicable syringes
Accept all brands of standard syringes:
10ml, 20ml, 50ml
Maximum power consumption
25W, can run for 4h after being recharged
 
Accumulated injection volume
0.1-999.9ml
 
Battery recharge
When the pump is connected to the AC power supply, turn on the power and the battery will automatically recharge (about 8-14 hours to recharge fully).
 
Injection accuracy
±3%
 
Feasible syringe
rate range
10ml:0.1ml/h-100ml/h;       
20ml:0,1ml/h-200ml/h;        50ml:0.1ml/h-500ml/h
Electromagnetic compatibility
Comply with the stipulation of IEC60601-1-2 :2006
 
Displayed Information
Rate, Accumulated injection volume, Syringe size, Battery capacity, Bed No., AC Power indicator, Battery charge indicato
Operating conditions
Environment temperature +5ºC~+40ºC,atmosphere pressure 86~106kPa, relative humidity ≤80%
 
Storage
Environment temperature -40ºC~+55ºC, atmosphere pressure 50~106kPa, relative humidity≤95%.
 
Alarm function
Over, NEAR, Occlusion, Low battery, Syringe disengaged, No AC power supply, Serious low battery
 
Classification
Class II, type BF
 
Bolus rate
The fastest rate of the size of syringe in use
 
Certification
CE0197
 
Power supply
50/60Hz AC100V~240V
 
Applied standards
IEC 60601-1:2005
IEC60601-2-24:1998
IEC60601-1-2:2006   ISO780-1997 ISO15223:2000     ISO7886-1:1993  
ISO/IEC:Guide No.37
<Consumer Goods Usage Instructions>
 
Battery
Rechargeable lithium polymer battery,
7.4V, 1600mA
 
  Fuse
T2AL250V