Product Main

Specifications

Ultrasonic thickness gauge (UM-2D)
 
Standard package:
UM-2D thickness gauge
Standard probe (PT-08)
Carrying case
Two 1.5V AA alkaline batteries
Couplant
Operating manual
Certification of quality
Packaging list
 
Operating principle:
Standard mode: P-E (pulse-echo) method with dual-element probes
Coating mode: E-E (echo-echo) method with dual-element probes
 
Probe zero adjustment:
Selectable: 1-point calibration: normal measurement
                 2-point calibration: curve surface measurement or other special application with a high precision
 
V-path correction: Automatic
 
Measuring range:
0.8mm to 300mm; 0.031in to 11.81in in standard mode
3.0mm to 18.0mm; 0.118in to 0.709in in through coating mode
Depending on probe, material and surface condition
 
Digital resolution:
H<4in : 0.001in /0.01in or 0.01mm/0.1mm
H≥4in : 0.01in or 0.1mm
 
Units:
Selectable: inch or mm
 
Display:
128×64 dot-matrix LCD screen (1.65in×2.24in ;42mm×57mm)
EL backlight and switchable contract ratio
Digit height: 0.54in or 13.75mm
 
Measurement update rate:
4 Hz in standard measurement mode
25 Hz in MIN capture mode
 
Material velocity range: 1,000 to 9,999 m/s; 0.039 to 0.394 inch/µs
 
Power:
Two 1.5V AA alkaline batteries
Warning with low battery voltage
 
Operating time:
Up to 200 hours (without backlight) with alkaline batteries, depending on operating mode
 
Auto shut-off: after 5 minutes of non-use
 
Operating temperature:
-10℃ to +50℃; +14℉ to +122℉
Specification to -20℃/-4℉ on request
 
Size:
5.86in×2.87in×1.25in; 149mm×73mm×32mm (H×W×D)
 
Weight: 200g; 7oz (including batteries)
 
Data logger capacity:
Up to 500 readings can be divided into a maximum of 5 files (user-selectable)
 
Optional accessories:
Attachable rubber case
Numerous special probes on request
Couplants and high-temperature couplant
4-step/6-step block for calibration
Data transmission cable
Application software
 
Operating language:  English and Chinese
 
Measuring error:
Up to 1in: ±0.002in or ±0.05mm
Above 1in: ±0.5%H
H is the height of the testing object
 
Repeatability: ±0.001in or ±0.025mm
 
Application software: UMVIEW
 
Interfaces:
Serial RS 232C port in combination with a communicative cable;
Data transfer parameters: 8 bits, 1 stop bit, 9600 baud, no parity.