Model number: ECG-923 (Three channel)
Dimension: 310mmX235mmX66mm
Packing: 375mmX295mmX195mm
Weight: 4.5kg(gross); 3.57kg(net)
Dimension: 310mmX235mmX66mm
Packing: 375mmX295mmX195mm
Weight: 4.5kg(gross); 3.57kg(net)
Three channel with interpretive
Build in rechargeable Li battery
Chinese/English language
Simultaneous 12 leads
320X240 graphic LCD, 5.7 inch screen
Paper size: 80mmX20/30m
Built in RS232 interface (USB is optional)
128 patients save/copy/communication
Specification
|
Items
|
specification
|
|
Lead
|
standard 12 leads
|
|
lead acquisition
|
synchronously 12 leads
|
|
Data acquisition
|
18bit/1000Hz
|
|
Record mode
|
Manual mode; AUTO1; AUTO2; AUTO3
|
|
Filter
|
AC, EMG, Drift Fliter
|
|
CMRR
|
≥ 100dB(with AC filter)
|
|
Input circuit
|
Floating; Protection circuit against Defibrillator effect
|
|
Input circuit current
|
≤ 0.1μ A
|
|
Input Impedance
|
≥ 2.5MΩ
|
|
Patient current leakage
|
<10μ A
|
|
Calibrating voltage
|
1mV± 2%
|
|
Voltage tolerance
|
± 500mV
|
|
Time constant
|
>3.2s
|
|
Frequency response
|
0.05 ~150Hz
|
|
Noise level
|
≤ 15μ Vp-p
|
|
Sensitivity valve
|
≤ 20μ V
|
|
Inter-Channel Interference
|
≤ 0.5mm
|
|
Paper speed
|
25, 50mm/s(± 3%)
|
|
Sensitivity
|
AUTO, 2.5, 5, 10, 20mm/mV
|
|
Recording mode
|
Thermal Array printing
|
|
Paper specification
|
80mmx30m, roll
|
|
LCD display
|
. 320x240 graphic LCD, display status& waveform.
|
|
Safety classification
|
IEC60601-1 class I, type CF
|
|
Power supply
|
AC : 110V/220V/110~230V, 50/60Hz± 1Hz; 40VA
|
|
DC : 14.4V(2000mAh) rechargeable Li-ion battery
|
|
|
Fuse
|
AC extend T500mA/125V ¢5x20mm
AC extend T400mA/250V ¢5x20mm
AC extend T2.0A/250V ¢5x20mm
|