CMS 6000B Multi-parameter Monitor
The following FDA Disclaimer is required for all eBay listing in Healthcare category and is included for REFERENCE:
" The sale of this item may be subject to regulation by the U. S. Food and Drug Administration and state and local regulatory agencies. If so, do not bid on this item unless you are an authorized purchaser. If the item is subject to FDA regulation, I will verify your status as an authorized purchaser of this item before shipping of the item"
" The sale of this item may be subject to regulation by the U. S. Food and Drug Administration and state and local regulatory agencies. If so, do not bid on this item unless you are an authorized purchaser. If the item is subject to FDA regulation, I will verify your status as an authorized purchaser of this item before shipping of the item"
Features:
8.4″ color TFT display.
Suitable for adult, pediatric and neonatal patient.
Basic parameters (ECG, SpO2, NIBP) in a durable case for bedside monitoring and transport.
Audible and visual alarms with adjustable alarm ranges.
Networkable with CONTEC central monitoring system.
Powerful data management and storage capacity.
Technical Parameters:
Temperature:
Resolution, 0.1℃.
Channel, Dual-channel.
Alarm range, 0-50℃.
ECG:
Lead type: 5-lead.
Input: RA; LA; RL; LL; V.
Sweep speed: 12.5mm/s, 25mm/s, 50mm/s.
Accuracy: ± 1bpm or ± 1%, whichever is greater.
Protection: Withstand 4000VAC/50Hz voltage in isolation against electrosurgical and defibrillation.
S-T detection: YES, Arrhythmia analysis: YES.
Alarm: YES, audible and visual alarm, alarm events recallable.
NIBP:
Method: Oscillometry.
Operation modes: Manual/Automatic/STAT.
Measurement unit: mmHg/KPa selectable.
Measurement types: Systolic, Diastolic, Mean.
Over-pressure protection: YES.
SPO2:
Measurement range: 0~100%.
Accuracy: ± 2digit (70~100%), 0~69% unspecified.
Respiration:
Method: RA-LL impedance.
Pediatric/Neonatal: 6-150rpm.
Alarm range: Adult: 6-120rpm.
Pediatric/Neonatal: 6-150rpm.
Apnea alarm: YES.
Standard Configuration:
ECG, NIBP, SpO2, RESP, 2-TEMP, Network.
Optional:
2-IBP, ETCO2, Thermal Recorder.
EtCO2
Measurement Mode: Side-stream or mainstream.
Measurement range: EtCO2: 0-99mmHg.
Accuracy: CO2 concentration.
Apnea alarm: YES.
IBP
Channel: 2 channels.
Pressure transducer: Sensitivity: 5( mV/V/mmHg).
Transducer sites: ART, PA, CVP, RAP, LAP, ICP.
Alarm range: -10~300mmHg.