Product Main

Specifications

ECG (ECG-100G), CE approval certification No. G1 10 02 50972 013.
 
Features
■ High-resolution thermal printing array system
■ Frequency response is as high as 150Hz
■ Capable of printing continuously on channel trace and annotations including lead mark and parameter such as sensitivity, paper speed and filter operation status
■ High resolution digital filter inhibits baseline drift without affecting ECGwaveform
■ One-touch operation
■ Concept of floating input circuit meets IEC safety standard
■ Rechargeable battery supporting more than 50 patient examinations
■ Lead Standard: 12 leads
 
Specifications
 
Input Circuit Floating; Protection against Defibrillator effect
Lead Standard 12 Leads, Lead change automatically
Patient Current Leakage < 10μ A
Input Impedance ≥ 10Μ Ω
Calibrating Voltage 1mV± 3%
A/D Conversion 12 bit
Frequency Response 0.05Hz~150Hz (IEC)
Time Constant > 3.2seconds
CMRR
> 80dB
> 100dB (with filter)
EMG Filter 35Hz (-3dB) / 25Hz (-3dB)
Sensitivity 1/2, 1, 2 (cm/mV), Conversion Deviation≤ 5%
Recording System Thermal-array
8 dots/mm ( vertical )
16 dots/mm ( horizontal, 25mm/sec )
Paper Speed 25, 50mm/s± 3%
Recording Paper 50mm, 20m, high-speed roll
Input Circuit Current ≤ 50nA
Safety Standard IEC class I, type CF
Power Supply
AC: 220V / 110V, 50/60Hz
DC: 12V rechargeable battery
Skin Voltage Tolerance ± 300mV
Noise Level < 15uVp-p
Dimensions 310mm× 230mm× 70mm
Dimension of Packing 400mm× 280mm× 160mm
Weight 2.75kg