50KW High Frequency X-ray unit  YSX0607
1)Microcomputer control x-ray radiography assembles: optional 600mA or 50KW.
2)Total 432 storage combination controlled by element such as kV mA S is applied radiography for the following filed three body position, fat middle and thin body type, automatic detect and display cod of mistake in system, send warning signal, detect accuracy and maintain.
3)X-ray tube capacity protection, anode starting-up protection and radiograph time-limit protection device semi-conduct controlled, fluoroscopy computed time device, x-ray tube over-hot protectorate provided, ensure x-ray tube run safely long time, improve accuracy of kV mA S.
4)Bulb type over remote-controlled diagnostic table and image intensifier television system are installed. They realize not only bright operation, but also supply convenience for science resonator suffer are reduced. Specifications of automatic spot-film are abundant. There are also automatic collimators for users convenience.
5)Applied in all kinds of hospital for radiography such as general body, beam filtration, and chest.
Department
Technology parameters
High voltage generator control table and high voltage transform assemble
Fluoroscopy
Tube currency
0.2-12.5mA continuous currency
Tube voltage
40-150KV continuous voltage
Radiography
Tube currency
10-630mA shift
Tube voltage
40-150KV continuous voltage
Exposure time
0.01-10s 23 shift
Power
Voltage
380V±10%
Frequency
50Hz±1Hz
Capacity
≥50KVA
Inner resistance
≤0.15Ω
X-ray assembly
Power
50KW
Thermal Capacity
230kHU
Double focal
0.6/1.2mm
Changed beam limited device
Diagnosis table YS09D
Automatic, adjusted field of view
CDG
Multi function radiography table
Moving table
Lengthways
±600mm
Electromagnetism locking
Across
±120mm
 
Mobile beam filtration grid
Grid density
28L/cm
Grid ratio
8:1
Focusing distance
100cm
Scope of moving
≥600mm
Radiography item
5”×7”,8”×10”,10”×12”,11”×14”,
12”×15”,14”×14”,14”×17”
Distance from surface to film
75mm
Height of bed
700mm
Moving pillar
±180°,per 90°orientation, moving lengthways ≥1500mm
Bulb tube
Rotate the cross arm±180,ascend $ descend 1100mm
CS15
Electro-motion diagnostic table
The table body rotation
-15°~0°~+90°
Focus to film distance
1100~1500mm,adjustable manual
360°foot board
The table to height
870mm
Stationary grid
Grid density
40L/cm
Grid ration
10:1
Focusing distance
120cm
Spot film device
Scope  of moving
Longitudinal moving:900mm;
Transverse moving:±120mm
Specification of scope film
Full :8”×10”,10”×12”,11”×14”,14”×14”
Two:  8”×10”,10”×12”
Three:11”×14”,14”×14”
Four: 11”×14”,14”×14”
Medical television system
configuration
9”x-ray image intensifier made by Toshiba, CCD camera,14” high resolution monitor
14”monitor
resolution
Horizontal center≥1000 TV line
Gray scale
≥10 grades
Video width
>12MHZ
Mode of scan
625 lines,5 field,Interlaced scanning
Geometry distortion
<2%
SNR
≥40db
Power supply
220V  50Hz  AC
9”x-ray image intensifier made by Toshiba
resolution
≥46LP/mm
Conversion modulus
34cd/m2/μGY/s
Contrast
10% area:30:1,Φ10mm area:18:1
CCD
Japan Camera
Specification
1/2”,anti-halation,CCIR752×586
Option
 
LSJ pillar radiography frame
The scope of ascend and descend
1450mm
The minimum distance from focal spot
380mm
Radiography item
8”×10” 10”×12” 11”×14”12”×15”
14”×14” 14”×17”
Fixed beam filtration role
Grid density
40L/cm
Grid radio
10:1
Focal distance
180cm