3.3kW High Frequency Medical X-ray Unit YSX0301
USE
General radiography,Filter-ray-filter radiograph
Unit Configuration
Control table,X-ray tube unit,Diagnostic table
Main characteristics
Completely microprocessor controlled, ergonomic control panel with back-lighting digital LCD, two point technique(KV/mAs), programmable anatomical technique, high-power or low-power selection.
High frequency x-ray monoblock.
Radiography table can move in length and breadth.
The radiography table, pillar and vibarating ray-filter are in a whole without top and bottom track.
MAIN TECHNICAL INDEXES
USE
General radiography,Filter-ray-filter radiograph
Unit Configuration
Control table,X-ray tube unit,Diagnostic table
Main characteristics
Completely microprocessor controlled, ergonomic control panel with back-lighting digital LCD, two point technique(KV/mAs), programmable anatomical technique, high-power or low-power selection.
High frequency x-ray monoblock.
Radiography table can move in length and breadth.
The radiography table, pillar and vibarating ray-filter are in a whole without top and bottom track.
MAIN TECHNICAL INDEXES
ITEM
|
CONTENTS
|
INDEX
|
Generator
|
Operating frequency
|
40kHZ
|
Ripple
|
<3%
|
|
Anode slope
|
Stationary 12°
|
|
Focus
|
1.5mm
|
|
Nominal anodic power
|
3.3KW
|
|
Monoblock thermal capacity
|
665kHZ
|
|
Maximum currents and mAs
|
70mA/320mAs
|
|
power supply
|
Voltage/Frequency
|
220V±10%
|
Maximum impulsive
|
18A
|
|
Inner resistance
|
0.4Ω
|
|
Radiograph
|
KV
|
40~110KV Be adjustable continuously
|
mAs
|
0.2~320mAs In 28 steps
|
|
Radiographic table
|
Length×width×height of bed surface
|
2000mm×665mm×700mm
|
Moving range of bed surface
|
640mm in Length,200mm in breadth
|
|
Ray-filter
|
Travel in length ≥500mm
|
|
Grid density N=28
|
||
Convergence distance f.=100cm
|
||
Grid ratio r=8
|
||
Column of X-ray tube unit
|
Moving Longitudinally along
|
1000mm or 1800mm
|
Radiographic table
|
||
Distance to friend as moving up and down (along column)
|
650~1800mm
|
|
Rotary around the center of cross arm
|
±180°
|
|
Rotary around the shaft line of X-ray annular tubes
|
-10°~60°~ +120°
|
|
Max size of cassette
|
356mm×432mm(14”×17”)
|
Note: We will not make another notice if the contour of product and technical parameters will be changed