20kW High Frequency X-ray Unit(YSX0310)
Use:
1. Applied in fluoroscopy, radiograph, beam filtration radiograph and Gastrointestinal point radiograph.
2. Dual-CPU control, higher accuracy, better anti-interference ability.
3. Using high-frequency inverter technology, X-ray high-energy and high degree of single color, short exposure time and clear images. Doctors and patients receive much lower doses of X-ray.
4. Computer closed-loop control. Stable, accurate, reproducible output parameters.
5. With the Organ process photographic function, the user can modify and store any information, easy to operate. X-ray tube protection function can extend the life of X-ray tube. With fault diagnosis functions, the fault code will display real-time and it is reliable and easy to maintain.
6. Without rail, the diagnosis electric bed days is easy to install and simple to operate7. The best choice for the hospital to replace the power-frequency X-ray machine
Main Technology parameters
Use:
1. Applied in fluoroscopy, radiograph, beam filtration radiograph and Gastrointestinal point radiograph.
2. Dual-CPU control, higher accuracy, better anti-interference ability.
3. Using high-frequency inverter technology, X-ray high-energy and high degree of single color, short exposure time and clear images. Doctors and patients receive much lower doses of X-ray.
4. Computer closed-loop control. Stable, accurate, reproducible output parameters.
5. With the Organ process photographic function, the user can modify and store any information, easy to operate. X-ray tube protection function can extend the life of X-ray tube. With fault diagnosis functions, the fault code will display real-time and it is reliable and easy to maintain.
6. Without rail, the diagnosis electric bed days is easy to install and simple to operate7. The best choice for the hospital to replace the power-frequency X-ray machine
Main Technology parameters
1. Power voltage
|
a.c 380±10%
|
2. Power frequency
|
50HZ±1HZ
|
3. Max tube voltage
|
125kv
|
Power capacity
|
≥600W
|
4. Standard output power:
|
20KW
|
5. High voltage putout frequency
|
30Khz
|
6. Tube currency
|
Radiograph 16-200 mA, Fluoroscopy 0.3-5mA
|
7. Tube voltage
|
Radiograph 40~125kv; Fluoroscopy 40-110 kv
|
8. Currency time
|
0.5~320mAs 40 shift
|
9. Exposure time
|
Radiograph 0.0025-6.4s;
|
continuous Fluoroscopy
|
|
10. Focal
|
small focal : 0.6mm2, big focal: 1.2mm2,
|
11. Rated rotate speed of anode
|
2800r/min
|
12. Thermal capacity
|
140KHU
|
13. Diagnostic table
|
1960×700×820, No rail,
|
Rotation of the table: 90°-0°-10°, Electromagnetic Lock
|
|
Moving lengthways 1500mm,moving sideways 740mm
|
|
14. Beam filtration device
|
fixed beam filtration grid N=40line/cm
|
15. Point-chip device
|
Whole-chip: 5”×7”/ 8”×10”/ 10”×12”/ 11”×14”
|
Half-chip: 5”×7”/ 8”×10”/ 10”×12”
|
|
Quarter of the chip: 11”×14”
|
|
16. Fluorescent screen
|
Effective size 340mm×340mm;
Distance between focus and screen: 830±10mm
|
Irradiation vision size: not less than 330mm×330mm
|
Frame: Single tube high frequency high voltage generator / Control table / The diagnosis electric bed / High frequency x-ray tube assemble / High voltage cable (2 ) / Beam filtration device